265 øpurge, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 416

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-265

ØPURGE

Page:

O, S

Stand:

M, A

Abbreviation:

ØP

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. ØPURGE .....

Function

Switch off purging.

Description

The

ØPURGE

instruction is used to switch off the inert gas purging for the

measuring vessel on the 747 VA stand switched on with

PURGE

. Purging

is used to remove the electrochemically active and hence interfering
oxygen in aqueous solutions. With the inert gas flow rate of ca. 20 L/h set
on the 747 VA Stand, a purging time of ca. 3...5 min usually suffices.

Comment

If the purging should be switched off automatically after a definite time,
the

TPURGE

instruction can also be used.

see also

PURGE, TPURGE, STIR, TSTIR, END