beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 514

background image

7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-15

Error message

Meaning/Causes

Rectification

Exit

EVAL - ERROR 7:
AC2: illegal baseline
scope

With measurement mode

AC2MODE

, only

whole

is

allowed for

Scope

Set

Scope

to

whole

E/C

EVAL - ERROR 8:
AC2: illegal baseline
type

With measurement mode

AC2MODE

, selection of

polynom

or

expon.

for the baseline

type is not admissible

Set baseline type to

AC2 rel.

,

AC2

abs.

or

linear

(identical to

AC2

abs.

)

E/C

EVAL - ERROR 9:
Wave: illegal baseline
scope

In the measurement mode

DCTMODE

, only

whole

is

admissible for

Scope

Set

Scope

to

whole

E/W

GRAPHICS - ERROR 1:
Display sector too
small

The selected display range is

too small

Expand display range (change dis-

play/plot parameters)

E/C

KEYBOARD - ERROR 1:
Keyboard hardware
defect

Keypad hardware faulty

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 2:
HSA is ON, RS module
has not set HSA off

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 3:
HSB is ON before data
is set

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 4:
HSA and HSB are ON
before data is set

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 5:
CNT is ON in spite of
write cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 6:
CNT is not ON in spite
of read cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 7:
HSB is not set to ON at
read cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 8:
HSA is not set to OFF
at read cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 9:
HSB is not set to OFF
at read cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 10:
HSA + HSB are not set
to OFF at read cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 11:
Keyboard interface,
data not equal

Initializing error

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 12:
EOS is not set to ON
after configuration

Initializing error

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 13:
HSA is not set to OFF
after write cycle

Handshake error in the RS232

module

Press Reset key

If error reappears, inform Metrohm

service

S

KEYBOARD - ERROR 14:
Unknown keyboard error

Non-specified keypad error

Press Reset key

If error reappears, inform Metrohm

service

S