7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-13

Error message

Meaning/Causes

Rectification

Exit

DIAGNOSIS - ERROR 21:
No dosimats available

Diagnostic fault; no Dosi-

mats/Dosinos attached

Attach Dosimat/Dosino to 746 VA Trace

Analyzer

E/C

DIAGNOSIS - ERROR 22:
Dosimats timeout,
awaiting busyline ....

Diagnostic fault; error in

Dosimat test

Wait a short time then press

(retry)

If the error reappears, press

(abort> and inform Metrohm service

r/a

DIAGNOSIS - ERROR 23:
Dosi. wayregister
read/write failed

Diagnostic fault; error in

Dosimat test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 24:
Data card not available

Diagnostic fault; no data card

inserted or write protection

enabled

Insert data card or disable write protec-

tion

E/C

DIAGNOSIS - ERROR 25:
Data card memory test
failed

Diagnostic fault; error in testing

the data card

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 26:
Receive text Ifc.2 <>
send text Ifc.1

Diagnostic fault; error in RS test

Press (abort)

Quit diagnostic routine with

Reset RS interfaces (press

Ifc. 1/2>

on ”VA TRACE ANA-

LYZER” page)

Restart RS test in diagnostic routine

r/a

DIAGNOSIS - ERROR 27:
Receive text Ifc.1 <>
send text Ifc.2

Diagnostic fault; error in RS test

Press (abort)

Quit diagnostic routine with

Reset RS interfaces (press

Ifc. 1/2>

on ”VA TRACE ANA-

LYZER” page)

Restart RS test in diagnostic routine

r/a

DIALOGUE - ERROR 1:
This operation takes
... minutes

Display of the time needed for

filling Dosimats/Dosinos (if a

low filling rate has been

selected)

Press (retry) to continue the

filling

Press (abort) if the filling time is

too long. Increase filling rate and

restart filling process

r/a

DIALOGUE - ERROR 2:
Not enough free lines
on page

Insufficient free lines for

transfer of solutions with

on the

”SOLUTIONS” page

Insufficient free lines for

transfer of standard formulae

solutions with

method>

on the

”CALCULATION” page

Delete solutions not needed on the

”SOLUTIONS” page

Delete formulae not needed on the

”CALCULATION” page

E/C

DIALOGUE - ERROR 3:
RS232 configuration
timed out

The RS232 configuration can

not be changed at present

Wait a short time then press

(retry) and try again

If the error reappears, press

(abort) and reset RS interface with the

softkey

on the ”VA

TRACE ANALYZER” page

r/a

DIALOGUE - ERROR 4:
RS232 interface still
busy

The RS232 interface is still

active

Wait a short time then press

(retry) and try again

If the error reappears, press

(abort) and reset RS interface with the

softkey

on the ”VA

TRACE ANALYZER” page

r/a

DIALOGUE - ERROR 5:
No measuring stand
found

No VA Stand found when

instrument switched on

Switch off 746 VA Trace Analyzer

Attach 747 VA Stand properly

Switch on 746 VA Trace Analyzer

E