3.12 Connection of devices via RS232 interfaces

746 VA Trace Analyzer / 747 VA Stand

3-55

3.12.3 Connection of a balance

The balances listed below can be attached to RS232 interface 2 of the 746 VA

Trace Analyzer. The settings of the RS232 parameters on the 746 VA Trace Ana-

lyzer depend on the settings of the balance,

Receive mode

must be set to

on

.

Balance

Cable

Settings on 746 VA Trace Analyzer

Sartorius MP8, MC1

6.2125.070

Balance type Sartorius

Mettler AB, AG (LC-RS25)

cable from Mettler

Balance type Mettler

Mettler AM, PM

ME 33995 from Mettler:

green lead to pin 2,

brown to 3, white to 7,

yellow to 20 of the 25-

pin connector

Balance type Mettler

Mettler with interface 016

cable from Mettler:

red lead to pin 3, white

to 7 of the 25-pin

connector

Balance type Mettler

Mettler with interface 011

or 012

6.2125.020

Balance type Mettler

Mettler AT

ME 33995 from Mettler:

green lead to pin 2,

brown to 3, white to 7,

yellow to 20 of the 25-

pin connector

Balance type Mettler AT

Precisa

Balances with RS232

interface

6.2125.080


Balance type Precisa

AND

ER-60, 120, 180, 182

FR-200, 300

FX-200, 300, 320 with

RS232 interface (OP-03)

6.2125.020

6.2125.020

6.2125.020


Balance type AND
Balance type AND
Balance type AND


A balance and a printer can be attached to RS232 interface 2 at the same time with

the aid of the 6.2125.030 Socket-outlet adapter. The balance must then be plugged

into the ”data in” port of the socket-outlet adapter.
The initial mass is transferred as a number with up to 6 digits, sign and decimal

point. Units and control characters sent by the balance are not transferred. With the

aid of a special input device supplied by the balance manufacturer, not only the ini-

tial mass but also sample identifications and methods can be put in from the bal-

ance. The addresses of the identification or method must be preselected on the in-

put device:

Balance

Method

Id#1

Id#2

Id#3

Sartorius

METH or 27

ID.1 or 26

ID.2 or 24

C-20 or 23

Mettler (AT)

D (Mthd)

C (ID#1)

B (ID#2)

A (c20)