3 connection of a balance – Metrohm 746 VA Trace Analyzer User Manual
Page 82

3.12 Connection of devices via RS232 interfaces
746 VA Trace Analyzer / 747 VA Stand
3-55
3.12.3 Connection of a balance
The balances listed below can be attached to RS232 interface 2 of the 746 VA
Trace Analyzer. The settings of the RS232 parameters on the 746 VA Trace Ana-
lyzer depend on the settings of the balance,
Receive mode
must be set to
on
.
Balance
Cable
Settings on 746 VA Trace Analyzer
Sartorius MP8, MC1
6.2125.070
Balance type Sartorius
Mettler AB, AG (LC-RS25)
cable from Mettler
Balance type Mettler
Mettler AM, PM
ME 33995 from Mettler:
green lead to pin 2,
brown to 3, white to 7,
yellow to 20 of the 25-
pin connector
Balance type Mettler
Mettler with interface 016
cable from Mettler:
red lead to pin 3, white
to 7 of the 25-pin
connector
Balance type Mettler
Mettler with interface 011
or 012
6.2125.020
Balance type Mettler
Mettler AT
ME 33995 from Mettler:
green lead to pin 2,
brown to 3, white to 7,
yellow to 20 of the 25-
pin connector
Balance type Mettler AT
Precisa
Balances with RS232
interface
6.2125.080
Balance type Precisa
AND
ER-60, 120, 180, 182
FR-200, 300
FX-200, 300, 320 with
RS232 interface (OP-03)
6.2125.020
6.2125.020
6.2125.020
Balance type AND
Balance type AND
Balance type AND
A balance and a printer can be attached to RS232 interface 2 at the same time with
the aid of the 6.2125.030 Socket-outlet adapter. The balance must then be plugged
into the ”data in” port of the socket-outlet adapter.
The initial mass is transferred as a number with up to 6 digits, sign and decimal
point. Units and control characters sent by the balance are not transferred. With the
aid of a special input device supplied by the balance manufacturer, not only the ini-
tial mass but also sample identifications and methods can be put in from the bal-
ance. The addresses of the identification or method must be preselected on the in-
put device:
Balance
Method
Id#1
Id#2
Id#3
Sartorius
METH or 27
ID.1 or 26
ID.2 or 24
C-20 or 23
Mettler (AT)
D (Mthd)
C (ID#1)
B (ID#2)
A (c20)