Metrohm 746 VA Trace Analyzer User Manual
Page 311

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-160
5.6.4
CALCULATION
CALCU-
LATION
:
;
This dialog page is employed for the entry of calculation formulae
and designations of the result quantities and units. Variables and
operators can be used for free definition of the "Final results" put
out on the "RESULTS" page.
Method: XXXXXXXX CALCULATION
max. 15 lines
Quantity Formula (R##, C##, A##) Res.unit Sig.dig.
ДДДДДДДДДДДД
ДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДД
ДДДДДДДД
ДДДДДДДД
ДДДДДДДД
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
............ ........................................ ........ ........
Modify
column
Modify
line
Delete
line
Insert
line
New
calcul.
Upd.from
method
Syntax
check
Softkeys
Meaning
Modify
column
Modify parameters column-by-column
Modify
line
Modify parameters line-by-line
Delete
line
Delete line
The line selected with the field cursor is deleted.
Insert
line
Insert line
A blank line is inserted above the line selected with the field cursor.
New
calcul.
Recalculation
The content and formula calculation from the evaluation results of the
substances (EV values) is restarted.
Upd.from
method
Transfer of standard formulae from method
For all substances defined in the method, standard formulae for the
mass concentration are entered and provide the results in g/L or g/g,
depending on the SMPL command and the sample size
S0
.
Sytax
check
Syntax test
The put in formulae are checked for proper syntax. Any errors are