Metrohm 746 VA Trace Analyzer User Manual
Page 332

5.7.2 Overview of the program instructions
746 VA Trace Analyzer / 747 VA Stand
5-181
•
Dispensing of standard addition solutions
Instruction Meaning
Page
Stand
ADD/M
Standard addition solution is already in measuring
vessel (MEAS)
O, S
M
ADD>M
Dispense standard addition solution into measuring
vessel (MEAS)
O, S
M, A
PADD>M
Dispense standard addition solution proportional to
the measured evaluation value into measuring
vessel
O, S
M, A
•
Dispensing of calibration solutions
(calibration curve and internal standard)
Instruction Meaning
Page
Stand
CSOLN>M
Dispense calibration solution into measuring vessel
(MEAS)
O, S
M, A
CSOLN/M
Calibration solution is already in measuring vessel
(MEAS)
O, S
M
CSOLN/F
Calibration solution is already in sample vessel
(FEED)
O, S
A
RADD>M
Dispense reference solution with internal standard
into measuring vessel (MEAS)
O, S
M, A
RADD>F
Dispense reference solution with internal standard
into sample vessel (FEED)
O, S
A
•
Instructions for 695 Autosampler
Instruction Meaning
Page
Stand
A>M
Transfer solution from auxiliary vessel (AUX) to
measuring vessel (MEAS)
O, S
A
F>M
Transfer solution from sample vessel (FEED) to
measuring vessel (MEAS)
O, S
A
F>A
Transfer solution from sample vessel (FEED) to
auxiliary vessel (AUX)
O, S
A
RINSE
Rinse measuring vessel (MEAS) and reset volume
bookkeeping to "0"
O, S
M, A
VMIX
Mix solution in specified rack position
O, S
A
VWASH
Additional rinsing of transfer tubing after solution
transfer to target vessel
O, S
A