5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-273

(REP

Page:

O, S

Stand:

M, A

Abbreviation:

(R

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. (REP .....

Function

Start of the loop for replications.

Description

The

(REP

instruction opens a loop for replications. The instructions

inserted in this loop are executed repeatedly. The loop must be closed
with the

REP)n

instruction.

Comments

A program can contain several REP loops, but they must not be
nested.

ADD, CAL and VAR loops are allowed within REP loops.

A REP loop is allowed within ADD, CAL and VAR loops.

The

EXIT

instruction can be inserted in REP loops for premature exit

from this loop.

Example

Basic structure of a standard addition with repetitions of the measure-
ment in the segment:

:

SMPL>M or SMPL/M

Sample addition to measuring vessel

(ADD

Loop start for standard additions

(REP

Loop start for replication

SEGMENT

Call-up of the segment (measurement)

REP)n

Loop end, n = number of replications

ADD>M, ADD/M or PADD>M

Constant or proportional standard addition

ADD)n

Loop end, n = number of standard additions

:

see also

REP)n, (ADD, ADD)n, (CAL, CAL)n, (VAR, VAR)n