273 (rep, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 424

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-273
(REP
Page:
O, S
Stand:
M, A
Abbreviation:
(R
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. (REP .....
Function
Start of the loop for replications.
Description
The
(REP
instruction opens a loop for replications. The instructions
inserted in this loop are executed repeatedly. The loop must be closed
with the
REP)n
instruction.
Comments
•
A program can contain several REP loops, but they must not be
nested.
•
ADD, CAL and VAR loops are allowed within REP loops.
•
A REP loop is allowed within ADD, CAL and VAR loops.
•
The
EXIT
instruction can be inserted in REP loops for premature exit
from this loop.
Example
Basic structure of a standard addition with repetitions of the measure-
ment in the segment:
:
SMPL>M or SMPL/M
Sample addition to measuring vessel
(ADD
Loop start for standard additions
(REP
Loop start for replication
SEGMENT
Call-up of the segment (measurement)
REP)n
Loop end, n = number of replications
ADD>M, ADD/M or PADD>M
Constant or proportional standard addition
ADD)n
Loop end, n = number of standard additions
:
see also
REP)n, (ADD, ADD)n, (CAL, CAL)n, (VAR, VAR)n