Metrohm 746 VA Trace Analyzer User Manual
Page 52

3.4 Multi-mode electrode (MME)
746 VA Trace Analyzer / 747 VA Stand
3-25
3.4.9
Cleaning the MME
If the mercury in the multi-mode electrode is contaminated and this leads to distur-
bances (see sections 7.4.2 and 7.4.3), the MME must be cleaned and refilled with
ultrapure mercury. Proceed as follows:
1 Remove multi-mode electrode from 747 VA Stand
•
Unscrew FEP tubing 51
51 and 59
59 from the MME, disconnect electrode
cable 37
37 from MME.
•
Take multi-mode electrode 38
38 out of measuring head 41
41 while gently
tapping the MME to knock off any mercury drops on the glass capillary
into the measuring vessel.
2 Remove old mercury
•
Place multi-mode electrode 38
38 in a horizontal position in the electrode
holder 114
114 (see Fig. 15-3). The slotted screw 101
101 is now at the top.
•
Unscrew slotted screw 101
101 using a suitable coin.
•
Carefully turn MME and empty mercury through the threaded opening
100
100 into a waste container placed in the drip pan 113
113. While doing so,
gently tap the glass capillary 109
109 and the MME to ensure that all mercury
flows out of the MME.
3 Dismantle MME
•
Unscrew retaining nut 108
108.
•
Take glass capillary 109
109 out of opening 104
104, the sealing ring 106
106 and the
locking ring 107
107 are removed at the same time. Remove these two parts
from the glass capillary 109
109.
•
Remove insert ring 105
105 from the MME.
•
Unscrew slotted screw 96
96 with a suitable coin in an anticlockwise direc-
tion from screw thread 98
98.
•
Screw locking ring 95
95 onto slotted screw 96
96 and pull out of the MME.
4 Dispose of old capillary
•
Connect filling tubing 115
115 with the tubing coupling 118
118 to the two gas
wash bottles 117
117 and the vacuum pump (see Fig. 14).
•
Insert glass capillary 109
109 (capillary end) in the filling cone 116
116 of the filling
tubing 115
115.
•
Remove mercury from the capillary with the vacuum pump.
5 Clean MME
•
Clean inner compartments of the MME, contact pin 102
102 and the screw
threads 98
98, 100
100 and 104
104 with a lint-free cloth.
•
Thoroughly rinse all inner compartments of the MME and the unscrewed
individual parts with dist. water and then dry with N
2
.
Do not use any organic solvents.