3.4 Multi-mode electrode (MME)

746 VA Trace Analyzer / 747 VA Stand

3-25

3.4.9

Cleaning the MME

If the mercury in the multi-mode electrode is contaminated and this leads to distur-

bances (see sections 7.4.2 and 7.4.3), the MME must be cleaned and refilled with

ultrapure mercury. Proceed as follows:

1 Remove multi-mode electrode from 747 VA Stand

Unscrew FEP tubing 51

51 and 59

59 from the MME, disconnect electrode

cable 37

37 from MME.

Take multi-mode electrode 38

38 out of measuring head 41

41 while gently

tapping the MME to knock off any mercury drops on the glass capillary

into the measuring vessel.

2 Remove old mercury

Place multi-mode electrode 38

38 in a horizontal position in the electrode

holder 114

114 (see Fig. 15-3). The slotted screw 101

101 is now at the top.

Unscrew slotted screw 101

101 using a suitable coin.

Carefully turn MME and empty mercury through the threaded opening

100

100 into a waste container placed in the drip pan 113

113. While doing so,

gently tap the glass capillary 109

109 and the MME to ensure that all mercury

flows out of the MME.

3 Dismantle MME

Unscrew retaining nut 108

108.

Take glass capillary 109

109 out of opening 104

104, the sealing ring 106

106 and the

locking ring 107

107 are removed at the same time. Remove these two parts

from the glass capillary 109

109.

Remove insert ring 105

105 from the MME.

Unscrew slotted screw 96

96 with a suitable coin in an anticlockwise direc-

tion from screw thread 98

98.

Screw locking ring 95

95 onto slotted screw 96

96 and pull out of the MME.

4 Dispose of old capillary

Connect filling tubing 115

115 with the tubing coupling 118

118 to the two gas

wash bottles 117

117 and the vacuum pump (see Fig. 14).

Insert glass capillary 109

109 (capillary end) in the filling cone 116

116 of the filling

tubing 115

115.

Remove mercury from the capillary with the vacuum pump.

5 Clean MME

Clean inner compartments of the MME, contact pin 102

102 and the screw

threads 98

98, 100

100 and 104

104 with a lint-free cloth.

Thoroughly rinse all inner compartments of the MME and the unscrewed

individual parts with dist. water and then dry with N

2

.

Do not use any organic solvents.