4.1 Preparations

746 VA Trace Analyzer / 747 VA Stand

4-3

6

Inserting reference electrode

Section 3.3.2

Inserting reference electrode in measuring head

and connecting

7

Inserting Pt auxiliary electrode

Section 3.3.2

Inserting Pt auxiliary electrode in measuring head

and connecting

8

Filling MME

Section 3.4

Filling multi-mode electrode (MME) with mercury,

mounting capillary and filling

9

Inserting MME

Section 3.3.2

Inserting multi-mode electrode (MME) in measuring

head and connecting

10 Filling measuring vessel

Section 3.3.2

Filling measuring vessel with ca. 20 mL dist. H

2

O and

inserting

11 Connecting inert gas

Section 3.3.3

4.1.3

Settings on the "VA TRACE ANALYZER" page

The settings necessary for startup of the 746 VA Trace Analyzer include setting the

date and time and contrast adjustment of the LCD. Proceed as follows:

1

Switch on instrument

Switch on the 746 VA Trace Analyzer with mains switch 11

11 at the rear of

the instrument. The instrument is initialized. If all is in order and the 747

VA Stand is attached properly, the dialog page "MONITORING" appears

after a few seconds.

The 746 VA Trace Analyzer has a total of 17 different dialog pages which can all be
selected directly using the yellow keys of the same name. The dialog keys are divided
into four groups and can also contain additional subpages. You will find an overview of
the dialog keys in section 5.1.3 and details of the individual pages in sections 5.3

…5.6.

2

Switch to dialog page "VA TRACE ANALYZER"

Press the yellow key. The "VA TRACE ANA-

LYZER" dialog page appears: