3.10 Connection of 700 Dosinos

746 VA Trace Analyzer / 747 VA Stand

3-45

3.10.4 Changing the Dosing unit

If the Dosing unit is changed when the Dosino is in the

dosing

posi-

tion, the Dosino can get destroyed! Therefore always make sure that

the Dosino is in the

exchange

position (see parameter

Position

on

dialog page "DOSIMATS") before changing the Dosing unit.

The Dosing unit mounted on the 700 Dosino can be changed only in the

exchange

position. Please proceed as follows:

1 Switch to exchange position

Select dialog page "DOSIMATS".

Select desired Dosino with the cursor keys and press softkey

position>

: The Dosino is switched to the exchange position and the pa-

rameter

Position

is set to

exchange

.

2 Change Dosing unit

Unscrew connection nipple of the PTFE tubing of the 4-way microtip 47

47

from connection 1 and change Dosing unit (procedure see 700 Instruc-

tions for use, section 2.3 and 2.4).

Screw connection nipple of the PTFE tubing of the 4-way microtip 47

47 onto

connection 1 of the new Dosing Unit mounted on the Dosino.

3 Initialize Dosino

Select dialog page "DOSIMATS".

Select desired Dosino with the cursor keys and press softkey

Dosino>

for initializing the Dosino. The cylinder is filled and excess solu-

tion is expelled to the measuring vessel. With standard accessories, all

tubings to the measuring vessel will have been filled at the end of the ini-

tializing procedure.