5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-190

(ADD

Page:

O, S

Stand:

M, A

Abbreviation:

(

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. (ADD .....

Function

Start of the loop for standard addition.

Description

The

(ADD

instruction opens a loop for standard additions. This loop

must include one of the dispensing instructions

ADD/M

,

ADD>M

or

PADD>M

for addition of the standard addition solution as well as a sweep or
segment call-up. Any other additional instructions can be incorporated.
The loop must be closed with the

ADD)n

instruction.

Comments

A program can contain several ADD loops, but they must not be
nested.

REP loops are allowed within ADD loops, an ADD loop can also be
included within a REP loop.

No CAL or VAR loops are allowed within ADD loops.

In the last execution of an ADD loop, the instructions

ADD>M

,

ADD/M

or

PADD>M

are no longer executed.

The

EXIT

instruction can be included within ADD loops for a

premature exit from this loop.

Example

A standard addition is normally programmed on the "OP.SEQUENCE"
page with the following basic structure:

:

SMPL>M or SMPL/M

Sample addition to measuring vessel

(ADD

Loop start for standard additions

(REP

Loop start for replication

SEGMENT

Call-up of the segment (measurement)

REP)n

Loop end, n = number of replications

ADD>M, ADD/M or PADD>M

Constant or proportional standard addition

ADD)n

Loop end, n = number of standard additions

:

see also

ADD)n

,

ADD/M

,

ADD>M

,

PADD>M

,

EXIT

,

Standard addition (section 5.6.3 and section 6.9.4)