7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-62

7.7.5

Assessment of the determination

To assess the recorded lead determination, the results printed out in the full report

for the mass concentration of lead

Mass conc.

and its total scatter

MC.dev.

are

used (example, see section 4.2). The limit values of these two results depend

greatly on the care taken in the preparation of the analysis solution and in the dis-

pensing of the standard addition solutions.

When the procedure is carried out properly and carefully, the following results

should be expected:

Accuracy =

95

… 105 % :

Mass conc. = 0.95

… 1.05 g/L

Scatter

≤≤

±± 3 %

:

MC.dev.

≤≤ ±± 0.03 g/L (±± 3 %)

The reproducibility can be determined by performing the

Testpb

method several

times and comparing the results obtained for the mass concentration

Mass conc.

.