beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 299

background image

5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-148

Parameter

Range; Default value / Meaning

Determinations with the aid of a calibration curve are normally pro-
grammed with the following basic structure:

:
D SMPL>M or SMPL/M

Sample addition to measuring vessel

C (CAL

Loop start for calibration

C CSOLN>M

Addition of calibration solution to measuring vessel

(REP

Loop start for replication

SEGMENT

Call-up of segment (with sweep)

ØMEAS

Switch off voltage (solution change)

REP)n

Loop end, n = number of replications

C RINSE

Resetting of the volume bookkeeping to 0

C CAL)n

Loop end, n = number of calibration solutions – 1

:

First, this program must be used to record the calibration curve. For
this,

Run mode = calibration

must be set on the "MONITORING"

page. The instructions marked by

D

are ignored. The calibration

solutions (max. 10) are measured in sequence and the coefficients of
the calibration curve determined. To store these permanently, the
method must then be copied from the working memory into the
method memory. This occurs automatically with

Auto.batch = on

; an

unsecured method with the same name in the method memory will be
overwritten. A switch is then made to

Run mode = determination

and

the sample measured. Instructions marked by

C

are ignored.

The result determination with the aid of a calibration curve saves time
compared with standard additions, but is reliable only

if the matrix of all samples and calibration solutions is identical or
has no influence on the measurement

if all measurement parameters (capillary, temperature, etc.) re-
main unchanged during measurements

if the accuracy of the results obtained is checked regularly with
the standard addition method.

With regard to optimum accuracy and deviation, a number of rules

must be observed with calibration curves (see also section 6.9.5):

Check linearity range

Check the linearity range of the calibration curve for each sub-
stance.

Working in the linear range

In determinations in the linear range, calibration must be per-
formed above all in the upper and lower parts of this range and
as large a number of replications as possible chosen.

Working in the non-linear range

In determinations in the non-linear range, as many calibration
solutions as possible with concentrations evenly distributed
throughout the entire range must be recorded.

Checking the offset

With curves, check with

linear

or

nonlin

whether the offset

Y.offset

may be neglected. The size of the offset provides in-

formation on a possible systematic error or blank value.