5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-286

SIMRESET

Page:

O, S

Stand:

M, A

Abbreviation:

SIMR

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SIMRESET .....

Function

Resetting the 3.496.8380 VA Simulator.

Description

The

SIMRESET

instruction is used to reset the current increases initiated

on the 3.496.8380 VA Simulator with the

SIMPULSE

instruction for the

simulation of standard additions.

Comments

The 3.496.8380 VA Simulator and the associated 3.496.5190
Connecting cable are available as an option. The VA Simulator gen-
erates 2 polarographic current waves with half-wave potentials of ca.
–540 mV and ca. –1070 mV in a defined voltage range. You will find
further details in the 3.496.8380-SB Circuit Description supplied with
the VA Simulator.

To ensure that the

SIMRESET

instruction functions properly, a line time

1 s must be entered.

Example

see

SIMPULSE

see also

SIMPULSE