Metrohm 746 VA Trace Analyzer User Manual
Page 538

7.4 Troubleshooting
746 VA Trace Analyzer / 747 VA Stand
7-39
Peak is in the highest
µ
A range
•
The sample volume is too large: reduce the volume. Carry out the analysis
again.
•
The enrichment period (for HMDE) is too long: Reduce the time.
•
If necessary use a SMDE or DME electrode instead of HMDE.
Double peak
•
Check MME. If necessary, change capillary (see section 3.4.8) or replace
sealing needle (see section 3.4.9).
•
Organic components interfere with the analysis? Carry out a UV digestion or
other suitable sample preparation.
•
If a second element is present at the same potential: add this element to the
sample and carry out the analysis again. If the second peak has become
higher then the second element is present.
•
Might it be possible to selectively mask this second element with a complex
former?
•
Electrolyte solution too old: make up a new one. Its working life with organic
additives may be as short as 1 day or less.
•
For Cu: work without chlorides in the electrolyte or increase the chloride con-
centration massively.
•
Try another measurement mode like AC1. If one substance is reversible and
the second one irreversible, only the reversible substance is detected by
AC1.
Standard addition peaks displaced
•
If HMDE is used potential displacements of more than 20
…
30 mV are often
normal and have to be accepted; particularly in adsorption voltammetry.
•
Standard solutions have been made too acidic.
•
Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-
ume.
No addition
•
Has the correct standard solution been used or is the concentration of the
solution too low: increase the volume of the standard addition or use a higher
concentration or reduce the sample amount accordingly.
•
If organic components are present: carry out a UV digestion or other suitable
sample preparation.
•
Concentration of the analyte is too high: dilute.
•
Addition solution with metal complexing solution (time reaction).