Metrohm 746 VA Trace Analyzer User Manual
Page 149

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-64
5
Switch to dialog page "DOCUMENTATION"
•
Press the yellow
MENTATION" appears for the automatic output of reports and curves.
The line entered here has the following meaning:
Curve Smth Sgmt:ASV VR:00
Printout of the smoothed segment curve
00 on the printer
•
If you do not want an automatic printout in the following determination,
you can either delete this line with the softkey
or clear
the output destination
Printer
for the line with the
•
Press the softkey
. The subpage of "DOCUMENTATION"
appears with the method comment describing, among other things, the
procedure used in the determination.
•
Finally press the
TATION".
4.6.4
Performing the determination
To perform the determination, the following steps are necessary:
1
Switch to dialog page "MONITORING"
•
Press the yellow
whose header shows the loaded method
ASV
appears.
2
Switch the program run mode to
determination
If the parameter
Run mode
is set to
calibration
, proceed as follows:
•
Use the cursor keys to select the field
Run mode
.
•
Press the space bar to enter the edit mode.
•
Press the softkey
and confirm this entry with the
key.
3
Entry of the sample volume
Enter the sample volume as follows:
•
Use the cursor keys to select the first field below the title
Sample
size/S0
.
•
Press the space bar to enter the edit mode.
•
Enter
10 mL
and confirm this entry with the