5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-255

Examples

Determination of cadmium and lead by stripping voltammetry on the 695
Autosampler (with predilution 1:10'000):

Method ASV-2 .mth OPERATION SEQUENCE

Title Determination of cadmium and lead

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 (PREP

2 SMPL/F V.fraction mL V.total L

3 DOS>A Soln.name Buffer V.add 4.950 mL

4 F>A addition FEED -> AUX V.add 0.050 mL

5 VMIX Cycles 2 V.mix 4.000 mL

Rack pos. SA

6 PREP)

7 DOS>M Soln.name Buffer V.add 9.900 mL

8 A>M addition AUX -> MEAS V.add 0.100 mL

9 >PREP

10 PURGE

11 STIR 600.0 Rot.speed 2000 /min

12 (ADD

13 (REP

14 PURGE

15 STIR 20.0 Rot.speed 2000 /min

16 ØPURGE

17 SEGMENT Segm.name CdPbASV

18 REP)2

19 ADD>M Soln.name PbCD V.add 0.020 mL

20 ADD)2

21 ØMEAS U.standby mV

22 RINSE Cycles 1 V.rinse 40.000 mL

23 RINSE Cycles 2 V.rinse mL

24 END

Method ASV-2.mth SEGMENT

CdPbASV

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 STIR 5.0 Rot.speed 2000 /min

2 HMDE Drop size 4 Meas.cell normal

3 DPMODE U.ampl 50 mV t.meas 20.0 ms

t.step 0.20 s t.pulse 40.0 ms

4 MEAS 60.0 U.meas -800 mV

5 ØSTIR 5.0

6 SWEEP 20.6 U.start -800 mV U.step 6 mV

U.end -200 mV Sweep rate 30 mV/s

7 END

.

In the sample preparation block

(PREP ... PREP)

, 0.05 mL sample from

the sample vessel (FEED) are transferred to the auxiliary vessel (AUX)
where they are mixed with 4.95 mL buffer. 0.1 mL of this solution are then
transferred to the measuring vessel, mixed with 9.9 mL buffer and
purged for 10 min with stirring. During this time, the PREP block for the
next sample is started. After purging, lead and cadmium in the segment

PbCdSV

are determined by stripping voltammetry with double standard

addition.

see also

PREP), >PREP