5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-289

SMPL/F

Page:

O, S

Stand:

A

Abbreviation:

SMP

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SMPL/F ..... V.fraction ........mL V.total .........L

Function

Sample volume has been added to the sample vessel (FEED) on the 695
Autosampler.

Parameter

Range; Default value / Meaning

V.fraction

0.010...50.000 mL; -

Fractional volume

With diluted samples, here the volume fraction of the total volume

V.total

which has been added to the sample vessel must be entered.

The following condition holds:

V.fraction

≤≤ V.total

V.total

0.10 mL...100000 L; -

Total volume

With diluted samples, here the total volume of sample and dilution
solution must be entered. The following condition holds:

V.total

≥≥ V.fraction

Description

The

SMPL/F

instruction defines the sample volume in the sample vessel

(FEED) on the 695 Autosampler.

With undiluted samples (

V.fraction

and

V.total

blank), the sample

volume is given either by

S0

(on entry of

S0

as a volume) or by

S0/S2

(on entry of

S0

as a weight,

S2

is the density in g/mL).

With diluted samples (

V.fraction > 0

and

V.total

≥≥ V.fraction

), the

sample volume is given by

V.fraction

, irrespective of the entry of

S0

.

You will find further details on the sample size in the description of the
parameter

Sample size/S0

in section 5.4.4.

Comments

The

SMPL/F

instruction can be used only in operation with the 695

Autosampler.

If no other dispensing operations are performed in the sample vessel
(FEED), an asterisk

*

can also be entered for

Sample size/S0

. The

sample volume is then defined by the subsequent transfer into the
auxiliary or measuring vessel.

see also

SMPL>M, SMPL/M