289 smpl/f, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 440

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-289
SMPL/F
Page:
O, S
Stand:
A
Abbreviation:
SMP
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SMPL/F ..... V.fraction ........mL V.total .........L
Function
Sample volume has been added to the sample vessel (FEED) on the 695
Autosampler.
Parameter
Range; Default value / Meaning
V.fraction
0.010...50.000 mL; -
Fractional volume
With diluted samples, here the volume fraction of the total volume
V.total
which has been added to the sample vessel must be entered.
The following condition holds:
V.fraction
≤≤ V.total
V.total
0.10 mL...100000 L; -
Total volume
With diluted samples, here the total volume of sample and dilution
solution must be entered. The following condition holds:
V.total
≥≥ V.fraction
Description
The
SMPL/F
instruction defines the sample volume in the sample vessel
(FEED) on the 695 Autosampler.
With undiluted samples (
V.fraction
and
V.total
blank), the sample
volume is given either by
S0
(on entry of
S0
as a volume) or by
S0/S2
(on entry of
S0
as a weight,
S2
is the density in g/mL).
With diluted samples (
V.fraction > 0
and
V.total
≥≥ V.fraction
), the
sample volume is given by
V.fraction
, irrespective of the entry of
S0
.
You will find further details on the sample size in the description of the
parameter
Sample size/S0
in section 5.4.4.
Comments
•
The
SMPL/F
instruction can be used only in operation with the 695
Autosampler.
•
If no other dispensing operations are performed in the sample vessel
(FEED), an asterisk
*
can also be entered for
Sample size/S0
. The
sample volume is then defined by the subsequent transfer into the
auxiliary or measuring vessel.
see also
SMPL>M, SMPL/M