5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-218

For permanent storage of the coefficients of the calibration curve
determined in the recording of the calibration solutions, the method
must then be copied from the working storage to the method memory
(this occurs automatically with

Auto.batch = on

).

Examples

Lead determination using a calibration curve with the

CSOLN>M

instruction (basic structure only):

Method Cal-3 .mth OPERATION SEQUENCE

Title Calibration curve with CSOLN>M

================================================================================

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

1 D SMPL>M V.fraction mL V.total L

2 C (CAL

3 C CSOLN>M addition into MEAS

4 (REP

5 SEGMENT Segm.name DPASV

6 REP)2

7 ØMEAS U.standby mV

8 RINSE Cycles 2 V.rinse mL

9 C CAL)3

10 END

.

Method Cal-3 SUBSTANCES

Pb - DPASV

================================================================================

Calibration 1993-11-16 09:42 Coefficients

–––––––––––––––––––––––––––––– –––––––––––––––––––––––––

Technique cal.crv. Y.offset

Curve type 0/linear Slope 0.0006743

Temperature 20.0 °C Nonlin.

Mean dev. 3.043e–10

Range min 2 ug/L # CSoln.name Mass conc.

Range max 30 ug/L –––––––––––––––––––––––––

M.conc./cm 5 ug/L 0 Pb-2.5 2.5 ug/L

1 Pb-5 5 ug/L

V0.CSoln. 10.000 mL 2 Pb-10 10 ug/L

3 Pb-25 25 ug/L

In the program run mode

calibration

, 4 calibration solutions each of 10

mL are measured in succession and used to determine the calibration
curve. In work with the manual Stand, the prompt for addition

CSOLN

"CSoln.name" "Mass conc." > M

appears with every calibration solution.

Following the measurement the prompt

Change measuring vessel

to

change the measuring vessel appears. Lead is then determined using
the measured calibration curve.

If this program is used with the 695 Autosampler, the 4 calibration
solutions must be entered on the "SAMPLES" page. They are then
fetched in turn from the positions on the sample rack specified here and
transferred to the measuring vessel. Following every measurement, the
measuring vessel is rinsed twice. Lead is then determined in the

determination

mode using the measured calibration curve.

see also

(CAL, CAL)n, CSOLN/F, CSOLN/M, RINSE,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)