5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-203

Determinations u sing the internal standard are normally programmed

on the OP.SEQUENCE page with the following basic structure:

:
C (CAL

Loop start for calibration

C CSOLN>M or CSOLN/M

Addition of calibration solution to

:

measuring vessel

C RADD>M

Addition of internal standard to

:

measuring vessel

C (REP

Loop start for replication

C SEGMENT Seg1

Call-up of the segment

:

(measurement of the calibration solution)

C REP)n

Loop end, n = number of replications

C ØMEAS

Switch off voltage (solution change)

C RINSE

Reset volume bookkeeping

C CAL)n

Loop end, n = number of calibration solutions – 1

:
D SMPL>M or SMPL/M

Sample addition to the measuring vessel

D RADD>M

Addition of internal standard to

:

measuring vessel

D (REP

Loop start for replication

D SEGMENT Seg1

Call-up of the segment

:

(measurement of the sample)

D REP)n

Loop end, n = number of replications

:

As in the previous example, with the same number of replications for
calibration and determination in the internal standards method the
illustrative program can be shortened as follows:

:

D SMPL>M or SMPL/M

Sample addition to measuring vessel

C (CAL

Loop start for calibration

C CSOLN>M or CSOLN/M

Addition of calibration solution

:

to measuring vessel

RADD>M

Addition of internal standard

:

to measuring vessel

(REP

Loop start for replication

SEGMENT Seg1

Call-up of the segment (measurement of the

:

calibration solution or sample)

REP)n

Loop end, n = number of replications

C ØMEAS

Switch off voltage (solution change)

C RINSE

Reset volume bookkeeping

C CAL)n

Loop end, n = number of calibration solutions – 1

:

see also

CAL)n, CSOLN/M, CSOLN>M, CSOLN/F, ØMEAS, RINSE, EXIT,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)