5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-203
Determinations u sing the internal standard are normally programmed
on the OP.SEQUENCE page with the following basic structure:
:
C (CAL
Loop start for calibration
C CSOLN>M or CSOLN/M
Addition of calibration solution to
:
measuring vessel
C RADD>M
Addition of internal standard to
:
measuring vessel
C (REP
Loop start for replication
C SEGMENT Seg1
Call-up of the segment
:
(measurement of the calibration solution)
C REP)n
Loop end, n = number of replications
C ØMEAS
Switch off voltage (solution change)
C RINSE
Reset volume bookkeeping
C CAL)n
Loop end, n = number of calibration solutions – 1
:
D SMPL>M or SMPL/M
Sample addition to the measuring vessel
D RADD>M
Addition of internal standard to
:
measuring vessel
D (REP
Loop start for replication
D SEGMENT Seg1
Call-up of the segment
:
(measurement of the sample)
D REP)n
Loop end, n = number of replications
:
As in the previous example, with the same number of replications for
calibration and determination in the internal standards method the
illustrative program can be shortened as follows:
:
D SMPL>M or SMPL/M
Sample addition to measuring vessel
C (CAL
Loop start for calibration
C CSOLN>M or CSOLN/M
Addition of calibration solution
:
to measuring vessel
RADD>M
Addition of internal standard
:
to measuring vessel
(REP
Loop start for replication
SEGMENT Seg1
Call-up of the segment (measurement of the
:
calibration solution or sample)
REP)n
Loop end, n = number of replications
C ØMEAS
Switch off voltage (solution change)
C RINSE
Reset volume bookkeeping
C CAL)n
Loop end, n = number of calibration solutions – 1
:
see also
CAL)n, CSOLN/M, CSOLN>M, CSOLN/F, ØMEAS, RINSE, EXIT,
Calibration curve (section 5.6.3 and section 6.9.5),
Internal standard (section 5.6.3 and section 6.9.6)