5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-150

Parameter

Range; Default value / Meaning

linear

nonlinear

Y.offset

EV

ρρ

eff

eff

Slope

0/linear

0/nonlinear

EV

ρρ

eff

eff

Slope

0

0

Temperature

0.0...99.9 °C; 25.0 °C

Measurement temperature

Measurement temperature for calibration and measurement.
Owing to the large temperature dependence of the measured values
(

2%/°C), it is advisable to work with the 6.1418.220 thermostatted

measuring vessel and to enter the set temperature here.

Range min

0.001...999.99 ng...kg/L;

Lower limit of working range

Here, a lower limit can be optionally defined for the working range with
the calibration curve. If a mass concentration

Mass conc.

less than

the value entered here is found for the substance, an error message is
put out and the remark

conc. out of range

displayed on the

"RESULTS" page.
The lower limit of the working range (in terms of the effective mass
concentration

ρ

eff

) is marked by a vertical line in the display and on the

plot of the calibration curve.

Range max

0.001...999.99 ng...kg/L;

Upper limit of working range

Here, an upper limit can be optionally defined for the working range
with the calibration curve. If a mass concentration

Mass conc.

greater

than the value entered here is found for the substance, an error
message is put out and the remark

conc. out of range

displayed on

the "RESULTS" page.
The upper limit of the working range (in terms of the effective mass
concentration

ρ

eff

) is marked by a vertical line in the display and on the

plot of the calibration curve.