5.3 Set Up

746 VA Trace Analyzer / 747 VA Stand

5-42

Parameter

Range; Default value / Meaning

Dil.settings

display only

Dilution settings

Title for the dilution parameters of the Dosimats/Dosinos. With the aid
of these parameters and the softkey

standard

solutions can be prepared. For this operation, the 6.5611.010 pipet-
ting equipment must have been installed at the Dosimat/Dosino and
connected to the remote interface of the 746 VA Trace Analyzer (see

section 3.10.6).

The initial volume

V.init std

needed for the dilution is automatically

calculated according to the formula

V.init std = V.end dilution * C.end dilution / C.init std

If the initial volume

V.init std

is smaller than 10

µL, the error mes-

sage

DOSIMAT ERROR 11: Illegal Dilution settings

appears.

We recommend to make only dilutions with an initial volume

V.init std

≥ 20 µL.

C.init std

0.001...999.9 ug...kg/L; 1.00 g/L

Initial standard concentration

Initial concentration of the standard used for the preparation of new
standard solutions.

The initial standard concentration

C.init std

must be

2 times the

desired end concentration

C.end dilution

.

C.end dilution

0.001...999.9 ng...g/L; 10.00 mg/L

End concentration

Desired end concentration of the new standard solution to be pre-
pared.

The end concentration

C.end dilution

must be

0.5 times the initial

standard concentration

C.init std

.

V.end dilution

0.1...999.9 mL; 10 mL

End volume

Desired end volume of the new standard solution to be prepared.