5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-259
PSAMODE
Page:
O, S
Stand:
M, A
Abbreviation:
PS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. PSAMODE .....
Function
Measurement mode for Potentiometric Stripping Analysis.
Application
Potentiometric stripping analysis is mainly used to determine substances
in an organic matrix with the aid of mercury film electrodes without prior
digestion.
Description
In the
PSAMODE
measurement mode for potentiometric stripping analysis,
the analytes are deposited at the working electrode with the constant
voltage
U.start
during a predetermined time
t.depos
and the voltage
U then measured for 6 s as a function of time. The nature of the
measurement is shown below using the RDE (mercury film electrode) as
an example.
Adjustable parameters:
PSWEEP
U.start
t.depos
Resolution
F (Hg film)
t
U
t
U.start
t.depos
t.meas = 6 s
In the
PSAMODE
measurement mode, the voltage measurement U vs t
(segment curves) is used to calculate the retention times t/U vs U
(substance curves). This results in peak-shaped curves which can be
evaluated automatically (see section 5.6.3). The peak voltage
U.peak
is
characteristic of the substance, the peak area
t.peak
is proportional to
its concentration.
t/U
U
U.peak
t.peak
Area (Hg film)