6 Data processing and evaluation

746 VA Trace Analyzer / 747 VA Stand

6-24

6.9.4

Standard addition

In the standard addition method, a known amount of the analyte is added once or

several times to the sample. The addition may be performed manually or automati-

cally and is programmed with the following instructions and parameters on the dia-

log page "OP.SEQUENCE" (see section 5.7.3):

(ADD...ADD)n

Loop for standard addition

n

Number of standard additions

ADD...

Constant standard addition

Soln.name

Name of the standard addition solution

V.add

Addition volume

PADD...

Proportional standard addition

Soln.name

Name of the standard addition solution

Prop.subst.

Name of the reference substance for

proportional standard addition

Prop.coeff.

Proportionality coefficient

V.add = EV: * Prop.coeff.

The remaining parameters needed for the standard addition are entered on the

dialog page "SUBSTANCES/Calibration" (see section 5.6.3). The most important of

these are:

Technique

Selection of the calibration technique

std.add

Standard addition

Curve type

Curve type for calculation of the standard addition

linear

Straight line of type

y = a + bx

nonlin

Curve of type

y = a + bx + cx

4

Soln.name

Name of the standard addition solution (must be identical

to the name specified in the ADD instruction)

Mass conc.

Mass concentration of the standard addition solution

In the standard addition, the following procedure is used to calculate the sought

mass concentration

ρ

eff,s

eff,s

of the sample (see also example in section 4.2):

1

Measurement of the sample solution
The sample solution with the unknown mass concentration

ρ

eff,s

eff,s

of the sam-

ple is measured one or more times. This gives:

EV

s

Evaluation quantity for sample

EV

s

.mean

Mean value of the evaluation quantity for sample

Std.dev.

Standard deviation of the individual value

EV

s

2

Measurement of the spiked sample solutions
The sample solution is spiked one or more times with a standard solution of

known mass concentration

ρ

std

std

. Each of these spiked solutions is measured

one or more times. This gives:

EV

n

Evaluation quantity for n

th

spiked sample

EV

n

.mean

Mean value of the evaluation quantities for

n

th

spiked sample

Std.dev.

Standard deviation of the n

th

individual value

EV

n