4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-26
2
Switch to dialog page "RESULTS/Evaluation"
•
Press the softkey
. The dialog page "RESULTS" with the sub-
page "Evaluations" appears showing an overview of the substance and
peak evaluations of the determination loaded in the working storage:
Method : TESTPb RESULTS
Determ.: 03201614 Evaluations
Substance : Pb Comments
Mass conc.: 1.032 g/L Mass : 103.2 ug ––––––––––––––––––
MC.dev. : 0.011 g/L (1.05%) Add.mass : 100 ug
Cal.dev. : - V0.sample: 100 uL
VR U/mV I/nA I.mean Std.dev. I.delta Comments
––– ––––– –––––– –––––– ––––––– ––––––– ––––––––––––––––––
00 -355 -195.9 -194.7 1.569
01 -355 -195.3
02 -355 -192.9
10 -356 -385.2 -385.5 0.4942 -190.8
11 -356 -386.1
12 -356 -385.3
20 -357 -568.1 -568.5 1.968 -182.9
21 -357 -566.7
22 -357 -570.6
Select
subst.
Suspend/
activate
New
calcul.
>>
Calcul.
>>
Calibr.
•
You will find a brief explanation of the result data shown here in the
description of the full report in section 4.2.6.
•
On this subpage you have the possibility to ignore undesired peak
results (e.g. spikes) for subsequent calculations. Proceed as follows:
−
Use the cursor keys to select the desired result by its VR num-
ber.
−
Press the softkey
. In addition to the VR
number, the marking
!
appears to show that this peak result will
be ignored in subsequent calculations (this marking can also be
canceled using the same softkey).
−
Press the softkey
. This initiates a new calculation
whose effects on the substance evaluation are immediately ap-
parent on this page.
3
Switch to dialog page "RESULTS/Calibration"
•
Press the softkey
. The dialog page "RESULTS" with the
subpage "Calibration" appears showing the results of the substance cali-
brations: