beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 328

background image

5.7.2 Overview of the program instructions

746 VA Trace Analyzer / 747 VA Stand

5-177

5.7.2

Overview of the program instructions

The following overview tables provide a brief introduction to the program instruc-

tions separated by category. You will find more detailed information and an expla-

nation of the parameters associated with the instruction in section 5.7.3.
The overview tables for the instructions contain not only the instruction and its

meaning, but also the two columns Page (dialog page) and Stand (VA Stand).

These show where the instruction is admissible. The meaning of the letters is as

follows:

O

Instruction admissible on dialog page "OP.SEQUENCE"

S

Instruction admissible on dialog page "SEGMENTS"

M

Instruction admissible with manual 747 VA Stand

A

Instruction admissible with automatic 747 VA Stand and

695 Autosampler

Electrode

Instruction Meaning

Page

Stand

DME

Switch on dropping mercury electrode

O, S

M, A

HMDE

Switch on hanging mercury drop electrode

O, S

M, A

SMDE

Switch on static mercury drop electrode

O, S

M, A

RDE

Switch on rotating disk electrode

O, S

M, A

Measurement mode

Instruction Meaning

Page

Stand

AC1MODE

Measurement mode for alternating current
voltammetry of the 1

st

harmonic

O, S

M, A

AC2MODE

Measurement mode for alternating current
voltammetry of the 2

nd

harmonic

O, S

M, A

CYCMODE

Measurement mode for cyclic voltammetry

O, S

M, A

DCTMODE

Measurement mode for direct current voltammetry
(Direct Current Tast)

O, S

M, A

DPMODE

Measurement mode for differential pulse
voltammetry

O, S

M, A

PSAMODE

Measurement mode for potentiometric stripping
analysis

O, S

M, A

SQWMODE

Measurement mode for square wave voltammetry

O, S

M, A