7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-60

7.7.2

Loading the

TESTPb.mth method

To load the

TESTPb.mth

method from the method memory to the working storage,

the following steps are necessary:

1 Switch on instrument

Switch on the 746 VA Trace Analyzer with mains switch 11

11 at the rear of

the instrument.

2 Switch to dialog page "METHODS"

Press the yellow key. The dialog page ”METHODS”

appears showing the methods stored in the instrument.

3 Selection of the TESTPb.mth method

Use the cursor keys to select the

TESTPb.mth

method, which then be-

comes black-backed.

4 Copying the TESTPb.mth method to the working storage

Press the softkey

. The softkey boar changes and presents the

destinations for copying the selected method.

Press the softkey

. The

TESTPb.mth

method is copied

to the working storage.

7.7.3

Preparing the sample solution

The sample solution for the lead determination is prepared as follows:

1 Addition of 20 mL dist. water

Add 20 mL dist. or deion. water to the measuring vessel.

2 Addition of 0.5 mL KCl 3 mol/L (6.2308.020)

Add 0.5 mL potassium chloride c(KCl) = 3 mol/L to the measuring ves-

sel.

3 Addition of 100

µµL Pb ion standard solution (6.2301.100)

Add 100

µ

L Pb ion standard solution

ρ(Pb) = 1 g/L to the measuring

vessel (this results in 100

µ

g Pb in the measuring vessel).