Metrohm 746 VA Trace Analyzer User Manual
Page 559

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-60
7.7.2
Loading the
TESTPb.mth method
To load the
TESTPb.mth
method from the method memory to the working storage,
the following steps are necessary:
1 Switch on instrument
•
Switch on the 746 VA Trace Analyzer with mains switch 11
11 at the rear of
the instrument.
2 Switch to dialog page "METHODS"
•
Press the yellow
appears showing the methods stored in the instrument.
3 Selection of the TESTPb.mth method
•
Use the cursor keys to select the
TESTPb.mth
method, which then be-
comes black-backed.
4 Copying the TESTPb.mth method to the working storage
•
Press the softkey
. The softkey boar changes and presents the
destinations for copying the selected method.
•
Press the softkey
. The
TESTPb.mth
method is copied
to the working storage.
7.7.3
Preparing the sample solution
The sample solution for the lead determination is prepared as follows:
1 Addition of 20 mL dist. water
•
Add 20 mL dist. or deion. water to the measuring vessel.
2 Addition of 0.5 mL KCl 3 mol/L (6.2308.020)
•
Add 0.5 mL potassium chloride c(KCl) = 3 mol/L to the measuring ves-
sel.
3 Addition of 100
µµL Pb ion standard solution (6.2301.100)
•
Add 100
µ
L Pb ion standard solution
ρ(Pb) = 1 g/L to the measuring
vessel (this results in 100
µ
g Pb in the measuring vessel).