7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-21
Error message
Meaning/Causes
Rectification
Exit
PRINTER - ERROR 1:
Printer hardware faulty
Printer hardware faulty
−
Press
−
If error reappears, press Reset key
−
If error reappears, start diagnostic
routine and test printer
r/a
PRINTER - ERROR 2:
Printer not reading
data
Printer does not read data
(interface fault)
−
Press
−
If error reappears, press Reset key
−
If error reappears, start diagnostic
routine and test printer
r/a
PRINTER - ERROR 3:
Printer is not ready
Printer is still busy
−
Wait and press
−
If error reappears, press Reset key
−
If error reappears, start diagnostic
routine and test printer
r/a
PRINTER - ERROR 4:
Printer interface, data
valid not set
Hardware fault at printer
interface (diagnostic fault)
−
Press
−
If error reappears, inform Metrohm
service
r/a
PRINTER - ERROR 5:
Printer interface, data
not equal
Hardware fault at printer
interface (diagnostic fault)
−
Press
−
If error reappears, inform Metrohm
service
r/a
PRINTER - ERROR 6:
No paper! Load paper,
please!
Printer has run out of paper
−
Load paper
−
Press
r/a
PRINTER - ERROR 7:
Unknown printer error
Non-specified printer fault
−
Press
−
If error reappears, press Reset key
−
If error reappears, inform Metrohm
service
r/a
RACKDEF - ERROR 1:
Too many definitions
for this rack type
(max. ...)
Too many positions entered for
the selected rack type
Delete redundant positions
E
RACKDEF - ERROR 2:
Line .., defined
location mismatches
with area
The specified position is not
admissible (restricted zone)
Check position and re-enter
E
RESULTS - ERROR 1:
No evaluation results
available
No results available for loaded
method
Start new measurement
E/C
RESULTS - ERROR 2:
Ambiguous sweep
identification per
sweep
Different measurement meth-
ods for the same substance
Check program (only one measurement
mode per substance is admissible)
E/C
RESULTS - ERROR 4:
No substance evaluation
available
No substance evaluation
available
Start new measurement
E/C
RS IFC.# - ERROR 1:
RS232 hardware faulty
RS232 hardware faulty
(# = 1 or 2)
−
Press Reset key
−
If error reappears, start diagnostic
routine and run RS test
r/a
RS IFC.# - ERROR 2:
HSA is ON, RS module
has not set HSA off
Handshake error in the RS232
module
(# = 1 or 2)
−
Press Reset key
−
If error reappears, start diagnostic
routine and run RS test
r/a
RS IFC.# - ERROR 3:
HSB is ON before data
is set
Handshake error in the RS232
module
(# = 1 or 2)
−
Press Reset key
−
If error reappears, start diagnostic
routine and run RS test
r/a
RS IFC.# - ERROR 4:
HSA and HSB are ON
before data is set
Handshake error in the RS232
module
(# = 1 or 2)
−
Press Reset key
−
If error reappears, start diagnostic
routine and run RS test
r/a