7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-21

Error message

Meaning/Causes

Rectification

Exit

PRINTER - ERROR 1:
Printer hardware faulty

Printer hardware faulty

Press (retry)

If error reappears, press Reset key

If error reappears, start diagnostic

routine and test printer

r/a

PRINTER - ERROR 2:
Printer not reading
data

Printer does not read data

(interface fault)

Press (retry)

If error reappears, press Reset key

If error reappears, start diagnostic

routine and test printer

r/a

PRINTER - ERROR 3:
Printer is not ready

Printer is still busy

Wait and press (retry)

If error reappears, press Reset key

If error reappears, start diagnostic

routine and test printer

r/a

PRINTER - ERROR 4:
Printer interface, data
valid not set

Hardware fault at printer

interface (diagnostic fault)

Press (retry)

If error reappears, inform Metrohm

service

r/a

PRINTER - ERROR 5:
Printer interface, data
not equal

Hardware fault at printer

interface (diagnostic fault)

Press (retry)

If error reappears, inform Metrohm

service

r/a

PRINTER - ERROR 6:
No paper! Load paper,
please!

Printer has run out of paper

Load paper

Press (retry)

r/a

PRINTER - ERROR 7:
Unknown printer error

Non-specified printer fault

Press (retry)

If error reappears, press Reset key

If error reappears, inform Metrohm

service

r/a

RACKDEF - ERROR 1:
Too many definitions
for this rack type
(max. ...)

Too many positions entered for

the selected rack type

Delete redundant positions

E

RACKDEF - ERROR 2:
Line .., defined
location mismatches
with area

The specified position is not

admissible (restricted zone)

Check position and re-enter

E

RESULTS - ERROR 1:
No evaluation results
available

No results available for loaded

method

Start new measurement

E/C

RESULTS - ERROR 2:
Ambiguous sweep
identification per
sweep

Different measurement meth-

ods for the same substance

Check program (only one measurement

mode per substance is admissible)

E/C

RESULTS - ERROR 4:
No substance evaluation
available

No substance evaluation

available

Start new measurement

E/C

RS IFC.# - ERROR 1:
RS232 hardware faulty

RS232 hardware faulty

(# = 1 or 2)

Press Reset key

If error reappears, start diagnostic

routine and run RS test

r/a

RS IFC.# - ERROR 2:
HSA is ON, RS module
has not set HSA off

Handshake error in the RS232

module

(# = 1 or 2)

Press Reset key

If error reappears, start diagnostic

routine and run RS test

r/a

RS IFC.# - ERROR 3:
HSB is ON before data
is set

Handshake error in the RS232

module

(# = 1 or 2)

Press Reset key

If error reappears, start diagnostic

routine and run RS test

r/a

RS IFC.# - ERROR 4:
HSA and HSB are ON
before data is set

Handshake error in the RS232

module

(# = 1 or 2)

Press Reset key

If error reappears, start diagnostic

routine and run RS test

r/a