212 condc, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 363

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-212
CONDC
Page:
O, S
Stand:
M, A
Abbreviation:
CO
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. CONDC U.start ........mV Rate 1 ........mV/s
U.end ........mV Rate 2 ........mV/s
Cycles ........
Function
Electrochemical regeneration of carbon electrodes.
Parameter
Range; Default value / Meaning
U.start
-3000...3000 mV; -1200 mV
Initial voltage of conditioning cycles
Initial voltage for every conditioning cycle.
U.end
-3000...3000 mV; -100 mV
Final voltage of conditioning cycles
Final voltage for every conditioning cycle.
Cycles
1...60000; 10
Number of conditioning cycles
This parameter defines the frequency of conditioning of the electrode.
Rate 1
1...10000 mV/s; 1000 mV/s
Rate from U.start to U.end
Sweep rate for the conditioning cycles form
U.start
to
U.end
.
Rate 2
5...50000 mV/s; 10000 mV/s
Rate from U.end to U.start
Sweep rate for the conditioning cycles form
U.end
to
U.start
.
Description
The
CONDC
instruction starts a freely selectable number of conditioning
cycles for the electrochemical regeneration of carbon electrodes (e.g.
glassy carbon).
Comments
Before the call-up of
CONDC
, the
RDE
electrode must be selected. While
is possible to run
CONDC
with other electrode types, this is not realistic.
see also
RDE