Metrohm 746 VA Trace Analyzer User Manual

Page 238

background image

5.5.2 RESULTS

746 VA Trace Analyzer / 747 VA Stand

5-87

Parameter

Range; Default value / Meaning

V0.sample

display only

Sample volume in measuring vessel (variable V0)

Display of the sample volume in the measuring vessel at the start of
measurement with numeric value and unit (L, mL,

µ

L, etc.).

The sample volume in the measuring vessel can be used in formulae
as

VO

for further calculations. It should be noted that this value is put

out in such cases in the base unit L without prefix (see section 5.6.4).
This field remains empty (

) if no calibration technique has been

selected (

Technique = none

).

The nature of the sample volume display depends on whether

Sample size/S0

has been entered as a volume or a weight (see

section 5.4.4),

work has been performed with undiluted (

V.fraction

= empty)

or diluted (

V.fraction

not empty) samples (see section 5.4.4),

only part of the sample quantity defined by

Sample size/S0

or

V.fraction

has been transferred to the measuring vessel on the

695 Autosampler (dilution factor f

AS

).

The dependency of

V0.sample

on these factors is shown in the

following Table:

S0 S2

V.fraction

V.total

V0.sample without

Autosampler

V0.sample with

Autosampler

L

S0

S0 *

f

AS

g

d

S0/S2

S0/S2 *

f

AS

L

V.fraction V.total

V.fraction

V.fraction *

f

AS

g

V.fraction V.total

V.fraction

V.fraction *

f

AS

You will find several numeric examples to clarify this table in the
explanation of the parameter

Sample size/S0

on the "SAMPLES"

page (see section 5.4.4).

Comments

display only

Comments

Display of comments if any type of error or difficulties appeared in the
determination of the mass concentration

ρ

s

of the substance.

The following comments are possible:

conc. out of range

The mass concentration

ρ

s

of the substance

Mass conc.

is outside the working range

Range min...Range max

defined on the

"SUBSTANCES" page.

dev.estimated

The total deviation

MC.dev.

could only be

estimated (see sections 6.9.5/6.9.6).

no cal.deviation

Calculation of the calibration deviation

Cal.dev.

is not possible as insufficient cali-

bration points are available (see sections

6.9.5/6.9.6).

no meas.deviation

The determination of the measurement de-
viation

Meas.dev.

is not possible as only 1

measured point is available (see section

6.9.4).