background image

4.6 Basic method for DPASV

746 VA Trace Analyzer / 747 VA Stand

4-63

2

Switch to dialog page "SEGMENT"

Press the yellow key. The dialog page "SEGMENT"

appears with the program for measured value acquisition:

Method: ASV SEGMENT

ASV

Instructions t/s Main parameters Auxiliary parameters

ДДДДДДДДДДДДД

ДДДДДДДДДДДДД ДДДДД

ДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

1 STIR 5.0 Rot.speed 2000 /min

2 HMDE Drop size 4 Meas.cell normal

3 DPMODE U.ampl 50 mV t.meas 20.0 ms

t.step 0.10 s t.pulse 40.0 ms

4 MEAS 90.0 U.meas -1150 mV

5 ØSTIR 5.0

6 SWEEP 21.9 U.start -1150 mV U.step 6 mV

U.end 150 mV Sweep rate 60 mV/s

7 ØMEAS U.standby mV

8 END

Modify

column

Modify

line

Delete

instr.

Insert

instr.

Select

segment

Delete

segment

Plot

param.

Segment

name

The sequence of the segment

ASV

is explained in brief in what follows:

Line 1

Stirring of the analysis solution for 5 s

Line 2

Switching on the HMDE (hanging mercury drop electrode)

Line 3

Selection of the measurement mode DP (differential pulse)
with voltage amplitude 50 mV, voltage step 0.1 s,
measurement time 20 ms and pulse time 40 ms

Line 4

Application of a measurement voltage of –1150 mV for
90 s for the enrichment at the hanging mercury drop

Line 5

Switching off the stirrer and stabilization of the analysis
solution for 5 s

Line 6

Voltage sweep from –1150 mV ... 150 mV with a voltage
step of 6 mV

Line 7

Switch off voltage and current measurement

Line 8

End of segment (return to main program)

3

Switch to dialog page "SUBSTANCES"

Press the yellow key. The dialog page "SUBSTAN-

CES" appears with the parameters for evaluation of substances. Only

one substance

Subst1

with standard parameters without an entry for the

characteristic voltage

U.verify

is defined.

Press the softkey

< >> >

. The subpage of "SUBSTANCES" appears with

the parameters for calibration. There is no calibration for the substance

Subst1

(

Technique = none

).

4

Switch to dialog page "CALCULATION"

Press the yellow key. The dialog page "CALCULA-

TION" appears for calculation of the final results. No formulae are en-

tered.