beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 251

background image

5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-100

Parameter

Range; Default value / Meaning

I.peak

display only

Peak height

Determined peak height in A.

I.peak

is shown for curves recorded with one of the following meas-

urement modes:

DPMODE, SQWMODE, AC1MODE
DCTMODE with SSWEEP (peaks)
CYCMODE with only 1 ramp

I

U

I.peak

I.wave

display only

Wave height

Determined wave height in A.

I.wave

is shown for curves recorded with one of the following meas-

urement mode DCTMODE and SWEEP (for waves).

I.wave

I

U

I.abs, I.rel

display only

Current evaluation quantity for AC2 curves

Determined current evaluation quantity in A.

I.abs

or

I.rel

is shown for curves recorded with the measurement

mode AC2MODE.

I.abs

Type = AC2 abs.

I

f

I

r

I.rel = I

f

+ I

r

I

U

Type = AC2 rel.

I

U

P.peak

display only

Peak area

Determined peak area in W.

P.peak

is shown for curves recorded with one of the following meas-

urement modes:

DPMODE, SQWMODE, AC1MODE
DCTMODE with SSWEEP (peaks)
CYCMODE with only 1 ramp

In an evaluation with

Scope = f.half

or

Scope = r.half

, the area up

to the peak maximum is shown here.