Metrohm 746 VA Trace Analyzer User Manual
Page 318

5.6.5 DOCUMENTATION
746 VA Trace Analyzer / 747 VA Stand
5-167
Parameter
Range; Default value / Meaning
Method
display only
Method name
Name of the method loaded in the working memory.
Auto form feed
no, yes; no
Automatic page advance
The automatic page advance before every report and curve printout
can be switched on or off here.
no
No automatic page advance, printing is continuous
and without regard to perforation.
yes
Automatic page advance before every report and
curve printout.
Auto error
printing
no, yes; no
Automatic error printing
The automatic printing of error messages during program execution
can be switched on or off here.
no
No automatic printing of error messages.
yes
The error messages occurring during program execu-
tion (status
* BUSY *
) are continuously printed out
(together with date and time of occurrence) on the
printer..
COPY
Report ...,Curve ...;
Selection of objects for automatic output
These lines can be used for entry of the objects (reports or curves)
which are put out automatically at the specified destination after every
determination. Each step in the entry is supported by softkeys which
offer the current selection possibilities for direct incorporation.
Reports
All reports with the exception of
MethSpc
and
Mpts
start with the
same preamble containing the following details:
METROHM
Name of manufacturer
746 VA TRACE ANALYZER
Instrument designation
(5.746.0101)
Program number
Determ.
Name of determination
Modified
Modification time (
no
= unchanged)
Sample table
Name of sample table (if sample data
transferred from sample table)
User
User name
Run
Run number
Date
Date of recording
Time
Time of recording
Pos.
Position number of sample on sample
rack (with 695 Autosampler)
Ident.1/S1
Sample identification 1 or S1
Ident.2/S2
Sample identification 2 or S2
Ident.3/S3
Sample identification 3 or S3
Method.call
Method call-up
Sample size/S0
Sample size