Metrohm 746 VA Trace Analyzer User Manual

Page 77

background image

3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-50

Put the pipetting handle into the waste beaker so that the pipetting tip is

at about the same level as the container with the solution that should be

pipetted and press the black button on the handle (if you press

at this time, you would have to add the solution manually).

If the pipetting instruction is carried out for the first time or if the previous

pipetting instruction has been terminated by (see below), the

message

Preparing Dosimat #X

appears. Press the black button on the

handle.

After rinsing of the pipetting tubing and formation of an air bubble, the

message

Aspirating standard

appears. Put the pipetting handle into

the desired solution with the pipetting tip immersed ca. 1

5 mm and

press the black button on the pipetting handle.

After aspiration of the solution the message

Dispensing solution

appears. Put the pipetting handle into the measuring vessel with the pi-

petting tip touching the wall and press the black button on the handle.

The desired solution is dispensed into the measuring vessel. Afterwards

the message

Rinsing tube

appears. Put the pipetting tip into the rinsing

quiver installed to the waste beaker and press the black button on the pi-

petting handle.

The pipetting tip is rinsed with dilution solution. Afterwards the message

Next operation

/

press to break

appears. If you press the black

button on the handle or , the Dosino will be ready for the next

dispensing instruction, if you press , the Dosino is switched to the

exchange position.