Metrohm 746 VA Trace Analyzer User Manual
Page 558

7.7 Standard operating procedure in the context of GLP/ISO 9001 guidelines
746 VA Trace Analyzer / 747 VA Stand
7-59
5 Testing the stirrer
•
Press the softkey
several times and check the on/off switch-
ing of the stirrer motor.
6 Copying the TEST747.mth method to the working storage
•
Press the yellow
•
Use the cursor keys to select the
TEST747.mth
method.
•
Press the softkey
and then
.
7 Entry of the sample volume
•
Press the yellow
•
Use the cursor keys to select the parameter
Sample size/S0
.
•
Press the
•
Enter <
*
> and press
8 Start program
•
Press the
•
Attach the electrode cable as described in the instructions on the screen
to the clamping screws of the dummy cell.
9 Assessment of the linearity
•
The first of the printed curves should satisfy the following conditions:
−
In principle, the plotted diagonal must be straight. Small deviations
are partly due to the resolution of the printer.
−
Current at voltage
– 200 mV:
– 1.6
µ
A ... – 2.4
µ
A*
−
Current at voltage
+200 mV:
+1.6
µ
A ... +2.4
µ
A*
* These tolerance values refer to illustrative scatters, that is to the differences
between individual instruments. The value determined for a specific instrument
shows less scatter. The deviation depends mainly on the temperature and should
be around 10%.
10 Assessment of peak measurement
•
The second curve plot should satisfy the following conditions:
−
A symmetrical, gaussian-shaped peak should be plotted. The
evaluation must provide a result for the peak voltage U and the peak
current I, which are printed out at the end of the curve.
−
Peak voltage U :
– 450 mV ... – 550 mV