7.7 Standard operating procedure in the context of GLP/ISO 9001 guidelines

746 VA Trace Analyzer / 747 VA Stand

7-59

5 Testing the stirrer

Press the softkey

several times and check the on/off switch-

ing of the stirrer motor.

6 Copying the TEST747.mth method to the working storage

Press the yellow key.

Use the cursor keys to select the

TEST747.mth

method.

Press the softkey

and then

.

7 Entry of the sample volume

Press the yellow key.

Use the cursor keys to select the parameter

Sample size/S0

.

Press the key.

Enter <

*

> and press .

8 Start program

Press the key.

Attach the electrode cable as described in the instructions on the screen

to the clamping screws of the dummy cell.

9 Assessment of the linearity

The first of the printed curves should satisfy the following conditions:

In principle, the plotted diagonal must be straight. Small deviations

are partly due to the resolution of the printer.

Current at voltage

– 200 mV:

– 1.6

µ

A ... – 2.4

µ

A*

Current at voltage

+200 mV:

+1.6

µ

A ... +2.4

µ

A*

* These tolerance values refer to illustrative scatters, that is to the differences

between individual instruments. The value determined for a specific instrument
shows less scatter. The deviation depends mainly on the temperature and should
be around 10%.

10 Assessment of peak measurement

The second curve plot should satisfy the following conditions:

A symmetrical, gaussian-shaped peak should be plotted. The

evaluation must provide a result for the peak voltage U and the peak

current I, which are printed out at the end of the curve.

Peak voltage U :

– 450 mV ... – 550 mV