8.1 746 VA Trace Analyzer

746 VA Trace Analyzer / 747 VA Stand

8-1

8 Technical data

Subject to changes !
The listed technical data apply to an ambient temperature of 25°C.

8.1

746 VA Trace Analyzer

Brief characterization

Compact control and data acquisition instrument for max. three 747 VA Stands,

a 695 Autosampler and up to four 685 Dosimats or 700 Dosinos with screen,

printer (optional), keypad and 2 bi-directional RS232C interfaces.
Circuitry based on SMD technology, 32-bit microprocessor, 16 MHz clock fre-

quency, 2 MBytes battery-backed SRAM.
Real-time multitasking operating system with dynamic memory management.

Current measurement techniques

DC

Direct Current

DP

Differential Pulse

SQW

Square Wave (10 ... 250 Hz)

AC1

Phase-sensitive Alternating Current 1

st

harmonic

(10 ... 250 Hz)

AC2

Phase-sensitive Alternating Current 2

nd

harmonic

(10 ... 250 Hz)

PSA

Potentiometric Stripping Analysis (chronopotentiometry)

CV

Cyclic Voltammetry (digital ramp, 0.1 mV/s ... 100 V/s)

Data acquisition

SWEEP

Current I as a function of the voltage U: I = f(U)
Voltage sweep for DC, DP, SQW, AC1, AC2

SSWEEP

Current I as a function of the voltage U: I = f(U)
Voltage sweep for DC (if peaks must be evaluated instead
of waves)

FSWEEP

Current I as a function of the voltage U: I = f(U)
Voltage sweep for DC, DP, SQW, AC1, AC2
(measurement always in operation mode ”fast”)

PSWEEP

Voltage U as a function of the time t: U = f(t)
Voltage sweep for PSA

TSWEEP

Current I as a function of the time t: I = f(t)
Current measurement at constant voltage for DC, DP,
SQW, AC1, AC2

RAMP

Current I as a function of the voltage U: I = f(U)
Voltage sweep for CV