beautypg.com

4 troubleshooting, General rules for va trace analysis -33, Problems with dme/smde -35 – Metrohm 746 VA Trace Analyzer User Manual

Page 532: Problems with hmde -37, Problems with rde -40

background image

7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-33

7.4

Troubleshooting

7.4.1

General rules for VA trace analysis

In order to get correct results in VA trace analysis, a number of general rules must

always be observed:

Chemicals and equipment

The purity of the reagents plays an important role in determining the re-

sults. Extremely pure chemicals should be used for determining lower

concentrations (see VA Application Note V-49).

Measuring vessel, electrodes and all other equipment in contact with the

sample solution must be clean and free of contamination substances.

Electrolytes

The pH during a determination plays an important role (e.g. for Zn, Cd, Pb,

Cu it should be approx. 4.5). Acetate, Ammonium acetate or PIPES buffer

are often used. For more information see the Application Bulletins.

The electrolyte must be sufficiently conductive and concentrated.

The purity of the electrolytes and the cleanliness of the reagent bottles is

very important.

The working life of the electrolytes is limited, particularly for organic addi-

tives (buffer substances, complex formers). It may be necessary to make

up fresh solutions every day.

Standard solutions

The standard solutions should be made acidic (approx. pH = 1

2) and

stored in plastic bottles.

Diluted standard solutions (ppb range) are very unstable and must be

freshly made. They must also be made sufficiently acidic.

The concentration of the standard solutions must be arranged so that a

volume between 20 and 500

µ

L has to be added.

Standard additions are recommended. The peak height after the last addi-

tion should be 2

5 times higher than the sample peak.

1000 ppm solutions (self-made or commercially available) are often used

as stock solutions. They are stable over long periods of time. Dilutions

have to be made with dilute acids.

Samples

The amount of sample depends on the concentration of the element to be

determined.

If the sample matrix is known, a better assessment of the analysis can be

made (organic components?).

A digestion must be carried out on contaminated samples and on sam-

ples where contamination is suspected (see Metrohm Monograph
«

Sample preparation for techniques in voltammetric trace analysis»).

A lot of errors are made during sampling and when storing the sample.

Caution and a critical approach are required.

The sample should have a good solubility in the electrolyte and be mixable

with it.