5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-186

The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type

linear

,

polynom

or

expon.

(see section 5.6.3).

I

U

U.peak

I.peak

Comments

The

AC1MODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

AC1MODE

measurement mode.

With the

AC1MODE

measurement mode, the potentiostat always

operates in the operating mode "fast".

SWEEP

and

FSWEEP

are thus

identical for this mode.

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

AC2MODE, SWEEP, DME, SMDE, HMDE, RDE

n = Meas.cycles

Σ

(I

n+

– I

n–

)

n

I

=