213 csoln/f, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 364

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-213
CSOLN/F
Page:
O, S
Stand:
A
Abbreviation:
CS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. CSOLN/F ..... solution added into FEED
Function
Calibration solution is located in the sample vessel (FEED) on the 695
Autosampler.
Description
The
CSOLN/F
instruction is used for the recording of calibration curves
with the techniques
cal.crv.
and
int.std.
using the 695 Auto-
sampler. Within a CAL loop, it defines that the calibration solution is
located in the sample vessel (FEED) on the sample rack of the 695
Autosampler.
In contrast to the
CSOLN>M
instruction,
CSOLN/F
is needed only when the
calibration solutions on the sample rack of the 695 Autosampler should
be pretreated before transfer to the measuring vessel (e.g. transfer to the
auxiliary vessel with the
F>A
instruction and subsequent dilution with the
DOS>A
instruction).
Comments
•
The
CSOLN/F
instruction can be used only in operation with the 695
Autosampler.
•
The number of calibration solutions is defined with
n
in the
CAL)n
instruction. A valid mass concentration
Mass conc.
must be entered
on the "SUBSTANCES/Calibration" page for each of these solutions.
•
In the
V0.CSoln
parameter on the "SUBSTANCES/Calibration" page,
the volume valid for all calibration solutions that has been added to
the sample vessel (FEED) is entered. If no other dispensing opera-
tions are performed in this vessel, an asterisk
*
can also be entered
for
V0.CSoln
.
•
If work is performed with
Auto.samples = on
, on the "SAMPLES"
page sample data lines must be entered in immediate succession for
all n calibration solutions with the following holding for their parame-
ters:
Pos.
Position number on the sample rack
(freely selectable)
S1...S3
Sample identifications (only entries of the last
calibration solution are used and stored)
Method.call
Name of the method +
.C
(for mode
calibration
)
Sample Size/S0
Entry of an asterisk
*
(value not used, but field must not be blank)
•
For permanent storage of the coefficients of the calibration curve
determined in the recording of the calibration solutions, the method
must then be copied from the working storage to the method memory
(this occurs automatically with
Auto.batch = on
).