5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-159

Parameter

Range; Default value / Meaning

Temperature

0.0...99.9 °C; 25.0 °C

Measurement temperature

Measurement temperature for calibration and measurement.

Owing to the large temperature dependence of the measured values
(

2%/°C), it is advisable to work with the 6.1418.220 thermostatted

measuring vessel and to enter the set temperature here.

Soln.name

8 characters;

Name of solution with internal standard

The name of the solution which contains the internal standard must be
identical with the name entered in the instruction

RADD>M

under

Soln.name

.

If the solution with the internal standard has to be added automati-
cally, this solution must be entered on the "SOLUTIONS" page and
assigned a position

Pos.

(see section 5.3.4). This solution entered on

the "SUBSTANCES" page can be automatically copied to the solution
list with the softkey

.

Mass conc.

0.001...999.99 ng...kg/L;

Mass concentration

ρρ

r

of solution with internal standard

If the solution with the internal standard has to be added automati-
cally, the mass concentration must also be entered on the
"SOLUTIONS" page (see section 5.3.4).