4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-12

The sequence of the segment

SegPb

is explained in brief in what follows:

Line 1

Switching on the DME (dropping mercury electrode)

Line 2

Selection of the measurement mode DP (differential pulse)
with voltage amplitude –50 mV, voltage step 0.4 s, meas-
urement time 20 ms and pulse time 40 ms

Line 3

Voltage sweep of –200 mV ... –550 mV with a
voltage step of 6 mV

Line 4

Switch off voltage, current measurement and DME

Line 5

End of the segment (return to main program)

3

Switch to dialog page "SUBSTANCES"

Press the yellow key. The dialog page "SUBSTAN-

CES" appears with the parameters for evaluation of the substance lead:

Method: TESTPb SUBSTANCES

Pb - SegPb

Recognition Display / Plot

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДД

U.verify -380 mV I.scale auto

U.tol (+/-) 40 mV U.div 50.00 mV/cm

U.width min 25 mV U.begin mV

U.width max 120 mV U.end mV

I.threshold 200 pA

Baseline Evaluation

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДД

Type linear Mode VA

Scope whole Quantity I.peak

dU.front auto Sign. digits 4

S.front auto

dU.rear auto

S.rear auto

Modify

column

Modify

line

Select

subst.

Link to

segment

Delete

subst.

Subst.

name

>>

The header of the dialog page "SUBSTANCES" contains the name of the substance
Pb and the segment SegPb linked with this substance. This is followed by the
parameters for substance recognition (Recognition), baseline calculation (Baseline),
evaluation quantity (evaluation) and curve representation (Display/Plot).
You will find further information regarding this dialog page in section 5.6.3.

The following is a brief explanation of the most important parameters:

U.verify

Characteristic voltage for lead: –380 mV

U.tol

Tolerance for characteristic voltage: only peaks with
a peak voltage of –380 V

±

40 mV are recognized

U.width min

Minimum peak width: only peaks with a
peak width > 25 mV will be recognized

U.width max

Maximum peak width: only peaks with a
peak width < 120 mV will be recognized

Type

Baseline type: linear (tangent)

Scope

Baseline range: entire curve