5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-80

Parameter

Range; Default value / Meaning

Ident.1/S1

12 characters;

Sample identification 1 / Variable S1

This column can be used for optional entry of either the alphanumeric
sample identification 1 or the numeric variable S1. The following rules
hold for the entry:

If the first character is not a number, the entire entry is interpreted
as alphanumeric sample identification 1 (e.g.

Sample 456

,

B-56-

234-199

).

If the first character is a number or a period (point), the entire
entry is interpreted as numeric variable S1 (e.g.

125.849

,

4.629e-

08

). Subsequent entry of letters or special characters is no longer

possible. If you wish to enter a sample identification starting with
a number, you must thus precede the number by a non-numeric
character (e.g.

#

).

The sample-specific variable

S1

can be used in formulae (see section

5.6.4).

Ident.2/S2

12 characters;

Sample identification 2 / Variable S2

This column can be used for optional entry of either the alphanumeric
sample identification 2 or the numeric variable S2. The same rules as
for

Ident.1/S1

apply to the entry.

If the sample size

Sample size/S0

is entered as a weight,

S2

must

be entered as a density in g/mL (without manual entry the default
value

1.000

is automatically entered).

The sample-specific variable

S2

can be used in formulae (see section

5.6.4). If

S2

has been entered as a density in g/mL, this numeric value

must be multiplied by the factor 1000 in formula calculations as all
other variables are based on the base units g or L.

Ident.3/S3

12 characters;

Sample identification 3 / Variable S3

This column can be used for optional entry of either the alphanumeric
sample identification 3 or the numeric variable S3. The same rules as
for

Ident.1/S1

apply to the entry.

The sample-specific variable

S3

can be used in formulae (see section

5.6.4).

Method.call

11 characters;

Method call

For automatic call-up of a method, method name, stand number and
program run mode can be entered here. The entry must have the
following format:

xxxxxxxx.SR

xxxxxxxx

=

Method name

S

=

Stand address (

1...3

)

R

=

Run mode (

C,D,T

)

(

Calibration/Determination/

Test

)

With this entry, all methods loaded in the method memory are dis-
played in the softkeys and can thus be taken over directly.