7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-25
Error message
Meaning/Causes
Rectification
Exit
SEGM.
Instruction not allowed
in block: line ##
There is an inadmissible
instruction in the specified
program line
Delete inadmissible instruction in the
segment
Esc
SEGM.
Missing part of prep.
(PREP, PREP), >PREP
One of the three instructions for
a PREP block missing
In program in the segment
,
check that all three instructions
(PREP
,
PREP)
and
>PREP
are present
Esc
SEGM.
Invalid with VARNI:
line ##
Inadmissible parameter for
parameter variation
Check parameter name for
Param.name
in the segment
and correct
Esc
SEGM.
Invalid with (CSWEEP:
line ##
Inadmissible instruction in the
CSWEEP block (only
RAMP
is
allowed)
Delete inadmissible instruction in the
CSWEEP block in the segment
Esc
SEGM.
Inconsist. value in
.....: line ##
Inadmissible value for the
specified parameter
Change parameter value in the segment
Esc
SEGM.
Value inconsist. with
electrode: t.step in
line ##
t.step
value for specified
electrode too low
−
Increase
t.step
or reduce
t.modul
−
Change electrode type
Esc
SEGM.
Missing value in line
##
Missing parameter value in the
specified program line
Entry parameter value in the segment
Esc
SEGM.
Not allowed data-
generating: line ##
Inadmissible data-generating
instruction in the specified
program line
Delete inadmissible instruction in the
segment
Esc
SEGM.
Not allowed in segment:
line ##
Inadmissible instruction in the
segment
Delete inadmissible instruction in the
segment
Esc
SEGM.
Incorrect use of (CAL
... RINSE sequence
If several CAL loops are used in
a program, every loop must
have a
RINSE
instruction
Enter
RINSE
instructions in the seg-
ment
Esc
SEGM.
Sweep range too small
Sweep range too small
(< 8 measured values)
Increase sweep range or decrease
U.step
or
t.step
Esc
SEGM.
Sweep range too large
Sweep range too large
(> 1000 measured values)
Decrease sweep range or increase
U.step
or
t.step
Esc
SEGM.
Missing t.step
t.step
value missing or is
<1 ms (possible only in
variation loops)
Enter variation loop for
t.step
in the
segment
so that
t.step
remains >1 ms
Esc
SEGM.
Conditioning cycle
range too small
Initial voltage
U.start
and
final voltage
U.end
of the
CONDC
instruction are identical
Change
U.start
and/or
U.end
in the
CONDC
instruction
Esc
SOLUTIONS - ERROR 1:
Mass conc. conflict on
line ##
The mass concentrations of
solutions with the same name
do not match
Check mass concentrations
Mass
conc.
of the solutions on the
”SOLUTIONS”, ”OP.SEQUENCE”,
”SEGMENTS” and ”SUBSTANCES/
Calibration” page and change
Esc
STORAGE - ERROR 1:
Destination file
secured
Secured file can not be deleted
or overwritten
Unsecure file before deleting or over-
writing with
E/C
STORAGE - ERROR 2:
Can not find source
The file to be copied is not
available at the source speci-
fied
Check source (working storage or
RS232 interfaces) to see whether the file
to be copied exists
E/C
STORAGE - ERROR 3:
Entry table full
Maximum number of files (256)
reached
Delete old files
E/C