Metrohm 746 VA Trace Analyzer User Manual
Page 534

7.4 Troubleshooting
746 VA Trace Analyzer / 747 VA Stand
7-35
7.4.2
Problems with DME/SMDE
Low background current or unstable baseline
•
The electrode drops irregularly: check MME.
Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary
(see section 3.4.8) or replace sealing needle (see section 3.4.9).
•
Check electrolyte concentration and pH of the solution.
•
Check initial potential and final potential of the analysis.
•
If the ion concentration in the solution is too high: dilute the electrolyte.
•
If the concentration to be determined is considerably lower than anticipated:
increase sample volume or change mode (e.g. HMDE).
•
Has the sample been degassed? Degassing with nitrogen for at least 5 min is
recommended, for alkaline solutions approx. 10 min is recommended.
•
Check tapping mechanism on VA Stand. If tapping strength is too weak, turn
slotted screw 71
71 on the valve block during operation in an anticclockwise di-
rection until a drop falls each time the tapper is triggered.
•
Is the gas pressure correctly set (1 bar)?
•
Is the reference electrode sufficiently full (inside and outside, see section
3.6)?
Curves look like a "starry night"
•
The electrode drops irregularly: check MME.
Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary
(see section 3.4.8) or replace sealing needle (see section 3.4.9).
•
Check contact between needle and capillary. If necessary, clean the MME
(see section 3.4.9).
•
Have stirring or degassing been switched off during the measurement?
•
If the electrode drops much too quickly: reduce
t.step
in program.
•
Check tapping mechanism on stand. If tapping strength is too weak, turn
slotted screw 71
71 on the valve block during operation in an anticclockwise di-
rection until a drop falls each time the tapper is triggered.
Standard addition curves are not reproducible
•
Check program in the instrument (stirring time, etc.).
•
Check MME, change capillary if necessary (see section 3.4.8).
•
The pipetting process was not correct: was the pipetting unit used properly?
•
Repeat the analysis again or try out automatic standard addition with Dosi-
mat, Dosino or 6.5611.010 Pipetting equipment.
•
Pipetting the standard solutions must be carried out by one and the same
person or with the same instrument or the same pipette.
•
Check and test pipetting process. When were the pipettes last calibrated
(GLP)?
•
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
•
Are the calibration solutions too old?
•
Would a calibration curve or the internal standard method be more suitable?