7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-35

7.4.2

Problems with DME/SMDE

Low background current or unstable baseline

The electrode drops irregularly: check MME.

Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary

(see section 3.4.8) or replace sealing needle (see section 3.4.9).

Check electrolyte concentration and pH of the solution.

Check initial potential and final potential of the analysis.

If the ion concentration in the solution is too high: dilute the electrolyte.

If the concentration to be determined is considerably lower than anticipated:

increase sample volume or change mode (e.g. HMDE).

Has the sample been degassed? Degassing with nitrogen for at least 5 min is

recommended, for alkaline solutions approx. 10 min is recommended.

Check tapping mechanism on VA Stand. If tapping strength is too weak, turn

slotted screw 71

71 on the valve block during operation in an anticclockwise di-

rection until a drop falls each time the tapper is triggered.

Is the gas pressure correctly set (1 bar)?

Is the reference electrode sufficiently full (inside and outside, see section

3.6)?

Curves look like a "starry night"

The electrode drops irregularly: check MME.

Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary

(see section 3.4.8) or replace sealing needle (see section 3.4.9).

Check contact between needle and capillary. If necessary, clean the MME

(see section 3.4.9).

Have stirring or degassing been switched off during the measurement?

If the electrode drops much too quickly: reduce

t.step

in program.

Check tapping mechanism on stand. If tapping strength is too weak, turn

slotted screw 71

71 on the valve block during operation in an anticclockwise di-

rection until a drop falls each time the tapper is triggered.

Standard addition curves are not reproducible

Check program in the instrument (stirring time, etc.).

Check MME, change capillary if necessary (see section 3.4.8).

The pipetting process was not correct: was the pipetting unit used properly?

Repeat the analysis again or try out automatic standard addition with Dosi-

mat, Dosino or 6.5611.010 Pipetting equipment.

Pipetting the standard solutions must be carried out by one and the same

person or with the same instrument or the same pipette.

Check and test pipetting process. When were the pipettes last calibrated

(GLP)?

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

Are the calibration solutions too old?

Would a calibration curve or the internal standard method be more suitable?