5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-307

VARNI

Page:

O, S

Stand:

M, A

Abbreviation:

VARN

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. VARNI ..... Param.name ............ Delta ........

Function

Variation of the next instruction (Variation of Next Instruction).

Parameter

Range; Default value / Meaning

Param.name

Meas.cycles, Modul.freq., Ph.angle, Prep.cycles, Rot.speed,
t.line, t.meas, t.pulse, t.step, U.ampl, U.end, U.meas,
U.start, U.step; t.line

Name of the parameter to be varied

The name entered here must be identical to the name of the parameter
which should be varied in the following instruction. The following
parameters can be varied:

Meas.cycles

Number of measurement cycles with measurement
modes

SQWMODE

,

AC1MODE

,

AC2MODE

.

Modul.freq.

Modulation frequency with measurement modes

SQWMODE

,

AC1MODE

,

AC2MODE

.

Ph.angle

Phase angle with measurement modes

AC1MODE

,

AC2MODE

.

Prep.cycles

Number of preparation cycles with measurement
modes

SQWMODE

,

AC1MODE

,

AC2MODE

.

Rot.speed

Rotational speed with

STIR

,

TSTIR

,

RDE

.

t.line

Line time for all instructions with line time.

t.meas

Measurement time with measurement modes

DCTMODE

,

DPMODE

,

SQWMODE

.

t.pulse

Pulse time with measurement mode

DPMODE

.

t.step

Time interval for voltage step with measurement
modes

DCTMODE

,

DPMODE

,

SQWMODE

,

AC1MODE

,

AC2MODE

.

U.ampl

Pulse amplitude with measurement modes

DPMODE

,

SQWMODE

,

AC1MODE

,

AC2MODE

(with

DPMODE

a varia-

tion is possible only in either the negative or the
positive region).

U.end

Final voltage for

SWEEP

instructions.

U.meas

Measurement voltage for

MEAS

instructions.

U.start

Initial voltage for

SWEEP

instructions.

U.step

Voltage step for

SWEEP

instructions.