5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-307
VARNI
Page:
O, S
Stand:
M, A
Abbreviation:
VARN
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. VARNI ..... Param.name ............ Delta ........
Function
Variation of the next instruction (Variation of Next Instruction).
Parameter
Range; Default value / Meaning
Param.name
Meas.cycles, Modul.freq., Ph.angle, Prep.cycles, Rot.speed,
t.line, t.meas, t.pulse, t.step, U.ampl, U.end, U.meas,
U.start, U.step; t.line
Name of the parameter to be varied
The name entered here must be identical to the name of the parameter
which should be varied in the following instruction. The following
parameters can be varied:
Meas.cycles
Number of measurement cycles with measurement
modes
SQWMODE
,
AC1MODE
,
AC2MODE
.
Modul.freq.
Modulation frequency with measurement modes
SQWMODE
,
AC1MODE
,
AC2MODE
.
Ph.angle
Phase angle with measurement modes
AC1MODE
,
AC2MODE
.
Prep.cycles
Number of preparation cycles with measurement
modes
SQWMODE
,
AC1MODE
,
AC2MODE
.
Rot.speed
Rotational speed with
STIR
,
TSTIR
,
RDE
.
t.line
Line time for all instructions with line time.
t.meas
Measurement time with measurement modes
DCTMODE
,
DPMODE
,
SQWMODE
.
t.pulse
Pulse time with measurement mode
DPMODE
.
t.step
Time interval for voltage step with measurement
modes
DCTMODE
,
DPMODE
,
SQWMODE
,
AC1MODE
,
AC2MODE
.
U.ampl
Pulse amplitude with measurement modes
DPMODE
,
SQWMODE
,
AC1MODE
,
AC2MODE
(with
DPMODE
a varia-
tion is possible only in either the negative or the
positive region).
U.end
Final voltage for
SWEEP
instructions.
U.meas
Measurement voltage for
MEAS
instructions.
U.start
Initial voltage for
SWEEP
instructions.
U.step
Voltage step for
SWEEP
instructions.