Metrohm 746 VA Trace Analyzer User Manual

Page 144

background image

4.5 Basic method for DP polarography

746 VA Trace Analyzer / 747 VA Stand

4-59

If you do not want an automatic printout in the following determination,

you can either delete this line with the softkey

or clear

the output destination

Printer

for the line with the key.

Press the softkey

. The subpage of "DOCUMENTATION"

appears with the method comment describing, among other things, the

procedure used in the determination.

Finally press the key to return to the main page of "DOCUMEN-

TATION".

4.5.4

Performing the determination

To perform the determination, the following steps are necessary:

1

Switch to dialog page "MONITORING"

Press the yellow key. The dialog page "MONITORING"

whose header shows the loaded method

POL

appears.

2

Switch the program run mode to

determination

If the parameter

Run mode

is set to

calibration

, proceed as follows:

Use the cursor keys to select the field

Run mode

.

Press the space bar to enter the edit mode.

Press the softkey

and confirm this entry with the

key.

3

Entry of the sample volume
Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample

size/S0

.

Press the space bar to enter the edit mode.

Enter

10 mL

and confirm this entry with the key.

4

Start the determination

Press the red key.

The status display in the header of the dialog page "MONITORING"

switches to

* BUSY *

, the start time of the measurement is entered un-

der

Determination

and the determination is given a standard name,

also formed from the start time. Further, during the measurement the

measurement window is superimposed which shows the current pro-

gram line being executed and the remaining line time.

Shortly after the start of the determination, the screen shows the prompt

to add 10 mL sample to the measuring vessel.

5

Addition of sample

Remove stopper 25

25 on 747 VA Stand and use a pipette to add 10 mL of

drinking water to the measuring vessel through the pipetting opening.

Press the key.

The screen shows the prompt to add 1 mL acetate buffer to the measur-

ing vessel.