7.6 Diagnosis

746 VA Trace Analyzer / 747 VA Stand

7-55

Attach electrode cable WE to clamping screw WE-D.

2 diagrams are printed out and must be assessed as follows:

Segment WE-L (DCTMODE):

In principle, the plotted diagonal must be straight. Small deviations

are partly due to the resolution of the printer.

Current at voltage

– 200 mV:

– 1.6

µ

A ... – 2.4

µ

A*

Current at voltage

+200 mV:

+1.6

µ

A ... +2.4

µ

A*

* These tolerance values refer to illustrative scatters, that is to the differences

between individual instruments. The value determined for a specific instrument
shows less scatter. The deviation depends mainly on the temperature and should
be around 10%.

Segment WE-D (AC1MODE):

A symmetrical, gaussian-shaped peak should be plotted. The

evaluation must provide a result for the peak voltage U and the peak

current I, which are printed out at the end of the curve.

Peak voltage U :

– 450 mV ... – 550 mV