4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-30

The drinking water sample should be acidified immediately after the

sample has been taken by adding 10

µ

L conc. hydrochloric acid per

10 mL drinking water.

Owing to the large temperature dependence of the measured values (

2%/°C), we

recommend you to use the 6.1418.220 Thermostatable measuring vessel with the

calibration curve method.

4.3.2

Loading the method

CALCRV.mth

To load the method

CALCRV.mth

from the method memory into the working stor-

age, the following steps are necessary:

1

Switch on instrument

Switch on the 746 VA Trace Analyzer with mains switch 11

11 at the rear of

the instrument. The instrument is initialized. If all is in order, the dialog

page "MONITORING" appears after a few seconds.

2

Switch to dialog page "METHODS"

Press the yellow key. The "METHODS" dialog page ap-

pears showing the methods stored in the instrument.

3

Selection of the method

CALCRV.mth

Use the cursor keys to select the method

CALCRV.mth

, which then

becomes black-backed.

4

Copying the method

CALCRV.mth to the working storage

Press the softkey

. The softkey bar changes and presents the

destinations for copying the selected method.

Press the softkey

. The method

CALCRV.mth

is

copied to the working storage.