7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-18

Error message

Meaning/Causes

Rectification

Exit

MEASURE - ERROR 17:
Different V0.CSoln.

Different volume for calibration

solutions

V0.CSoln.

with

several substances with

calibration mode

cal.crv.

/int.std.

in same segment

Set

V0.CSoln.

on the ”SUBSTANCES/

Calibration” page equal for all sub-

stances

E/S

MEASURE - ERROR 18:
Different mass conc.

Mass concentrations on the

”SUBSTANCES/Calibration”

and ”SOLUTIONS” pages do

not agree

Set mass concentrations equal on both

pages

E/S

MEASURE - ERROR 19:
Missing mass
concentration
(SUBSTANCES)

The mass concentration

Mass conc.

is missing for

ADD and CSOLN instructions

No solution for ADD instruc-

tion specified

Enter

Mass conc.

on the

”SUBSTANCES/Calibration” page

Enter solution name

Soln.name

in

ADD instruction

E/S

MEASURE - ERROR 20:
Illegal instruction for
Autosampler

The

SMPL/M

and

CSOLN/M

instructions are not allowed

with the 695 Autosampler

Correct SMPL or CSOLN instructions on

the ”OP.SEQUENCE” or ”SEGMENTS”

page

E/C

MEASURE - ERROR 21:
V.sample max too small
for dosing

The maximum volume

V.sample max

of the sample

vessels on the sample rack is

too small for the volume to be

dispensed

Lower volume

V.add

for dispensing or

transfer instruction

E/S

MEASURE - ERROR 22:
Unknown volume in FEED
for dosing

The entry of the asterisk in

Sample size/S0

is here not

admissible in conjunction

with the

SMPL/F

instruction

The entry of the asterisk in

V0.CSoln.

is here not ad-

missible in conjunction with

the

CSOLN/F

instruction

Enter volume or mass in

Sample

size/S0

Enter volume in

V0.Csoln.

E/S

MEASURE - ERROR 23:
Illegal * in
calibration volume
V0.CSoln.

The entry of the asterisk in

V0.CSoln.

is not admissible in

operation without the 695

Autosampler

Enter volume in

V0.CSoln.

E/S

MEASURE - ERROR 24:
Illegal variable S2

No numeric value in

S2

. If

S0

has been entered as the mass,

the density must be entered in

S2

in g/mL

Enter density in g/mL in

S2

on the

”SAMPLES” page

E/N

MEASURE - ERROR 25:
Undefined electrode

Undefined electrode type

(electrode type must be

entered before the measure-

ment mode)

Enter instruction for the electrode type

before that for the measurement mode

on the ”OP.SEQUENCE” or

”SEGMENTS” page

E/S

MEASURE - ERROR 26:
Undefined measurement
mode

Undefined measurement

mode (measurement mode

must be entered before the

sweep)

Wrong measurement mode

(e.g.

CYCMODE

with

SWEEP

)

Enter instruction for the measurement

mode before that for the sweep on

the ”OP.SEQUENCE” or

”SEGMENTS” page

Enter correct measurement mode on

the ”OP.SEQUENCE” or

”SEGMENTS” page

E/S

MEASURE - ERROR 27:
Too many measuring
points

The maximum number of 1000

measured points for a sweep

was exceeded

Reduce number of measured points by

shortening the sweep (

U.start/U.end

)

or increase the voltage step

U.step

E/S

MEASURE - ERROR 28:
Missing (CSWEEP

Command for start of the cyclic

loop missing

Enter instruction

(CSWEEP

on the

”OP.SEQUENCE” or ”SEGMENTS”

page

E/S

MEASURE - ERROR 29:
Missing instruction

Program instruction necessary

for the procedure missing

Check program instructions on the

”OP.SEQUENCE” or ”SEGMENTS”

page

E/S