5.5.1 MONITORING

746 VA Trace Analyzer / 747 VA Stand

5-75

Parameter

Range; Default value / Meaning

Run

display only

Run number

This counter is incremented by one on completion of every determi-
nation.

The run number is reset when the instrument is switched on and on
reset to 0.

The run number is put out with all result reports and curve printouts.

Stand address

1, 2, 3; 1

Stand address

Address number of the 747 VA Stand set using rotary switch

35

35

(see

section 3.3.1).

Maximum three 747 VA Stands can be attached to the 746 VA Trace
Analyzer, but only one can be active at the same time. The parameter

Stand address

is used to switch between the different stands.

For subsequent measurements, the set stand address remains in
force until it is changed by manual entry or by automatic method call-
up with specification of the stand address (1, 2, 3).

Auto.samples

XXX of 160

on, off; off

Automatic transfer of sample data

The automatic transfer of the samples entered on the "SAMPLES"
page can be switched on or off here.

on

On start of a determination, the data in the top line on
the "SAMPLES" page are transferred to the 1

st

sample

data line on the "MONITORING" page and used for the
determination.

off

On start of a determination, the current data in the
sample data lines on the "MONITORING" page are
used.

In addition to the input field, the number of remaining sample data
lines on the "SAMPLES" page is displayed (max. 160).

Auto.batch

off, on, reset; off

Automatic series determination

The automatic processing of sample series for the specified number
of samples (see below) can be switched on or off here.

on

On start of the first determination with , fur-
ther determinations are performed automatically and
the sample counter incremented by 1 each time until
the specified number of samples has been reached (or
until a stop using the or
keys if

*

is set).

The

type

of transfer of sample data depends on the

setting of the parameter

Auto.samples

(see also

1

st

sample data line

).